Title: Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy

Abstract

Purpose: Internal 1.5-mm fiducial markers were used in real-time tumor-tracking radiotherapy (RT) for lung cancer. The fixation rate of the markers using the bronchial insertion technique, reliability of the setup using markers around the target volume, dislocation of the markers after real-time tumor-tracking RT, and long-term toxicity of marker insertion were investigated. Methods and Materials: Between July 2000 and April 2004, 154 gold markers were inserted into 57 patients with peripheral lung cancer. The distances between the implanted markers in 198 measurements in 71 setups in 11 patients were measured using two sets of orthogonal diagnostic X-ray images of the real-time tumor-tracking RT system. The distance between the markers and the chest wall was also measured in a transaxial CT image on 186 occasions in 48 patients during treatment planning and during follow-up. The median treatment time was 6 days (range, 4-14 days). Results: In 115 (75%) of the 154 inserted markers, the gold marker was detected throughout the treatment period. In 122 markers detected at CT planning, 115 (94%) were detected until the end of treatment. The variation in the distances between the implanted markers was within {+-}2 mm in 95% and {+-}1 mm in 80% during treatment. Themore » variation in the distances between the implanted markers was >2 mm in at least one direction in 9% of the setups for which reexamination with a CT scan was indicated. The fixation rate in the left upper lobe was lower than in the other lobes. A statistically significant relationship was found between a shorter distance between the markers and the chest wall and the fixation rate, suggesting that the markers in the smaller bronchial lumens fixed better than those in the larger lumens. A learning curve among the endoscopists was suggested in the fixation rate. The distance between the markers and the chest wall changed significantly within a median of 44 days (range, 16-181 days) after treatment. Conclusion: The fixation of markers into the bronchial tree was useful for the setup for peripheral lung cancer and had an accuracy of {+-}2 mm during the 1-2-week treatment period. The relationship between the markers and tumor can change significantly after 2 weeks, suggesting that adaptive four-dimensional RT is required.« less

Authors:
 [1];  [2];  [3];  [3];  [3];  [3];  [4];  [5];  [6];  [3];  [1]
  1. First Department of Medicine, Hokkaido University School of Medicine, Sapporo (Japan)
  2. First Department of Medicine, Hokkaido University School of Medicine, Sapporo (Japan). E-mail: kyamazak@med.hokudai.ac.jp
  3. Department of Radiology, Hokkaido University School of Medicine, Sapporo (Japan)
  4. Department of Medicine, Iwamizawa Municipal Hospital, Iwamizawa (Japan)
  5. Department of Respiratory Diseases, Sapporo City General Hospital, Sapporo (Japan)
  6. Department of Medical Oncology, Hokkaido University Graduate School of Medicine, Sapporo (Japan)
Publication Date:
OSTI Identifier:
20788237
Resource Type:
Journal Article
Resource Relation:
Journal Name: International Journal of Radiation Oncology, Biology and Physics; Journal Volume: 63; Journal Issue: 5; Other Information: DOI: 10.1016/j.ijrobp.2005.04.024; PII: S0360-3016(05)00713-3; Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; ACCURACY; CAT SCANNING; CHEST; GOLD; IMAGES; LUNGS; NEOPLASMS; PATIENTS; PLANNING; RADIOTHERAPY; RELIABILITY; TOXICITY; X RADIATION

Citation Formats

Imura, Mikado, Yamazaki, Koichi, Shirato, Hiroki, Onimaru, Rikiya, Fujino, Masaharu, Shimizu, Shinichi, Harada, Toshiyuki, Ogura, Shigeaki, Dosaka-Akita, Hirotoshi, Miyasaka, Kazuo, and Nishimura, Masaharu. Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy. United States: N. p., 2005. Web. doi:10.1016/J.IJROBP.2005.0.
Imura, Mikado, Yamazaki, Koichi, Shirato, Hiroki, Onimaru, Rikiya, Fujino, Masaharu, Shimizu, Shinichi, Harada, Toshiyuki, Ogura, Shigeaki, Dosaka-Akita, Hirotoshi, Miyasaka, Kazuo, & Nishimura, Masaharu. Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy. United States. doi:10.1016/J.IJROBP.2005.0.
Imura, Mikado, Yamazaki, Koichi, Shirato, Hiroki, Onimaru, Rikiya, Fujino, Masaharu, Shimizu, Shinichi, Harada, Toshiyuki, Ogura, Shigeaki, Dosaka-Akita, Hirotoshi, Miyasaka, Kazuo, and Nishimura, Masaharu. Thu . "Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy". United States. doi:10.1016/J.IJROBP.2005.0.
@article{osti_20788237,
title = {Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy},
author = {Imura, Mikado and Yamazaki, Koichi and Shirato, Hiroki and Onimaru, Rikiya and Fujino, Masaharu and Shimizu, Shinichi and Harada, Toshiyuki and Ogura, Shigeaki and Dosaka-Akita, Hirotoshi and Miyasaka, Kazuo and Nishimura, Masaharu},
abstractNote = {Purpose: Internal 1.5-mm fiducial markers were used in real-time tumor-tracking radiotherapy (RT) for lung cancer. The fixation rate of the markers using the bronchial insertion technique, reliability of the setup using markers around the target volume, dislocation of the markers after real-time tumor-tracking RT, and long-term toxicity of marker insertion were investigated. Methods and Materials: Between July 2000 and April 2004, 154 gold markers were inserted into 57 patients with peripheral lung cancer. The distances between the implanted markers in 198 measurements in 71 setups in 11 patients were measured using two sets of orthogonal diagnostic X-ray images of the real-time tumor-tracking RT system. The distance between the markers and the chest wall was also measured in a transaxial CT image on 186 occasions in 48 patients during treatment planning and during follow-up. The median treatment time was 6 days (range, 4-14 days). Results: In 115 (75%) of the 154 inserted markers, the gold marker was detected throughout the treatment period. In 122 markers detected at CT planning, 115 (94%) were detected until the end of treatment. The variation in the distances between the implanted markers was within {+-}2 mm in 95% and {+-}1 mm in 80% during treatment. The variation in the distances between the implanted markers was >2 mm in at least one direction in 9% of the setups for which reexamination with a CT scan was indicated. The fixation rate in the left upper lobe was lower than in the other lobes. A statistically significant relationship was found between a shorter distance between the markers and the chest wall and the fixation rate, suggesting that the markers in the smaller bronchial lumens fixed better than those in the larger lumens. A learning curve among the endoscopists was suggested in the fixation rate. The distance between the markers and the chest wall changed significantly within a median of 44 days (range, 16-181 days) after treatment. Conclusion: The fixation of markers into the bronchial tree was useful for the setup for peripheral lung cancer and had an accuracy of {+-}2 mm during the 1-2-week treatment period. The relationship between the markers and tumor can change significantly after 2 weeks, suggesting that adaptive four-dimensional RT is required.},
doi = {10.1016/J.IJROBP.2005.0},
journal = {International Journal of Radiation Oncology, Biology and Physics},
number = 5,
volume = 63,
place = {United States},
year = {Thu Dec 01 00:00:00 EST 2005},
month = {Thu Dec 01 00:00:00 EST 2005}
}
  • Purpose: To evaluate implanted markers as a surrogate for tumor-based setup during image-guided lung cancer radiotherapy with audiovisual biofeedback. Methods and Materials: Seven patients with locally advanced non-small-cell lung cancer were implanted bronchoscopically with gold coils. Markers, tumor, and a reference bony structure (vertebra) were contoured for all 10 phases of the four-dimensional respiration-correlated fan-beam computed tomography and weekly four-dimensional cone-beam computed tomography. Results: The systematic/random interfractional marker-to-tumor centroid displacements were 2/3, 2/2, and 3/3 mm in the x (lateral), y (anterior-posterior), and z (superior-inferior) directions, respectively. The systematic/random interfractional marker-to-bone displacements were 2/3, 2/3, and 2/3 mm in themore » x, y, and z directions, respectively. The systematic/random tumor-to-bone displacements were 2/3, 2/4, and 4/4 mm in the x, y, and z directions, respectively. All displacements changed significantly over time (p < 0.0001). Conclusions: Although marker-based image guidance may decrease the risk for geometric miss compared with bony anatomy-based positioning, the observed displacements between markers and tumor centroids indicate the need for repeated soft tissue imaging, particularly in situations with large tumor volume change and large initial marker-to-tumor centroid distance.« less
  • Purpose: Internal fiducial gold markers, safely inserted with bronchoscopy, have been used in real-time tumor-tracking radiotherapy for lung cancer. We investigated the histopathologic findings at several points after the insertion of the gold markers. Methods and Materials: Sixteen gold markers were inserted for preoperative marking in 7 patients who subsequently underwent partial resection of tumors by video-assisted thoracoscopic surgery within 7 days. Results: Fibrotic changes and hyperplasia of type 2 pneumocytes around the markers were seen 5 or 7 days after insertion, and fibrin exudation without fibrosis was detected 1 or 2 days after insertion. Conclusions: Because fibroblastic changes startmore » approximately 5 days after gold marker insertion, real-time tumor-tracking radiotherapy should be started >5 days after gold marker insertion.« less
  • Purpose: To compare the retention rates of two types of implanted fiducial markers for stereotactic ablative radiotherapy (SABR) of pulmonary tumors, smooth cylindrical gold 'seed' markers ('seeds') and platinum endovascular embolization coils ('coils'), and to compare the complication rates associated with the respective implantation procedures. Methods and Materials: We retrospectively analyzed the retention of percutaneously implanted markers in 54 consecutive patients between January 2004 and June 2009. A total of 270 markers (129 seeds, 141 coils) were implanted in or around 60 pulmonary tumors over 59 procedures. Markers were implanted using a percutaneous approach under computed tomography (CT) guidance. Postimplantationmore » and follow-up imaging studies were analyzed to score marker retention relative to the number of markers implanted. Markers remaining near the tumor were scored as retained. Markers in a distant location (e.g., pleural space) were scored as lost. CT imaging artifacts near markers were quantified on radiation therapy planning scans. Results: Immediately after implantation, 140 of 141 coils (99.3%) were retained, compared to 110 of 129 seeds (85.3%); the difference was highly significant (p < 0.0001). Of the total number of lost markers, 45% were reported lost during implantation, but 55% were lost immediately afterwards. No additional markers were lost on longer-term follow-up. Implanted lesions were peripherally located for both seeds (mean distance, 0.33 cm from pleural surface) and coils (0.34 cm) (p = 0.96). Incidences of all pneumothorax (including asymptomatic) and pneumothorax requiring chest tube placement were lower in implantation of coils (23% and 3%, respectively) vs. seeds (54% and 29%, respectively; p = 0.02 and 0.01). The degree of CT artifact was similar between marker types. Conclusions: Retention of CT-guided percutaneously implanted coils is significantly better than that of seed markers. Furthermore, implanting coils is at least as safe as implanting seeds. Using coils should permit implantation of fewer markers and require fewer repeat implantation procedures owing to lost markers.« less
  • Purpose: The purpose of the present study was to perform an analysis of the changes in the dynamic seroma cavity based on fiducial markers in early stage breast cancer patients treated with accelerated partial breast irradiation (APBI) using three-dimensional conformal external beam radiotherapy (3D-CRT). Methods: A prospective, single arm trial was designed to investigate the utility of gold fiducial markers in image guided APBI using 3D-CRT. At the time of lumpectomy, four to six suture-type gold fiducial markers were sutured to the walls of the cavity. Patients were treated with a fractionation scheme consisting of 15 fractions with a fractionalmore » dose of 333 cGy. Treatment design and planning followed NSABP/RTOG B-39 guidelines. During radiation treatment, daily kV imaging was performed and the markers were localized and tracked. The change in distance between fiducial markers was analyzed based on the planning CT and daily kV images. Results: Thirty-four patients were simulated at an average of 28 days after surgery, and started the treatment on an average of 39 days after surgery. The average intermarker distance (AiMD) between fiducial markers was strongly correlated to seroma volume. The average reduction in AiMD was 19.1% (range 0.0%-41.4%) and 10.8% (range 0.0%-35.6%) for all the patients between simulation and completion of radiotherapy, and between simulation and beginning of radiotherapy, respectively. The change of AiMD fits an exponential function with a half-life of seroma shrinkage. The average half-life for seroma shrinkage was 15 days. After accounting for the reduction which started to occur after surgery through CT simulation and treatment, radiation was found to have minimal impact on the distance change over the treatment course. Conclusions: Using the marker distance change as a surrogate for seroma volume, it appears that the seroma cavity experiences an exponential reduction in size. The change in seroma size has implications in the size of the CTV, PTV, and percent of normal breast tissue irradiated when using 3D-CRT.« less
  • Purpose: To determine whether setup margins can be reduced using cone-beam computed tomography (CBCT) to localize tumor in conventionally fractionated radiotherapy for lung tumors. Methods and Materials: A total of 22 lung cancer patients were treated with curative intent with conventionally fractionated radiotherapy using daily image guidance with CBCT. Of these, 13 lung cancer patients had sufficient CBCT scans for analysis (389 CBCT scans). The patients underwent treatment simulation in the BodyFix immobilization system using four-dimensional CT to account for respiratory motion. Daily alignment was first done according to skin tattoos, followed by CBCT. All 389 CBCT scans were retrospectivelymore » registered to the planning CT scans using automated soft-tissue and bony registration; the resulting couch shifts in three dimensions were recorded. Results: The daily alignment to skin tattoos with no image guidance resulted in systematic ({sigma}) and random ({sigma}) errors of 3.2-5.6 mm and 2.0-3.5 mm, respectively. The margin required to account for the setup error introduced by aligning to skin tattoos with no image guidance was approximately 1-1.6 cm. The difference in the couch shifts obtained from the bone and soft-tissue registration resulted in systematic ({sigma}) and random ({sigma}) errors of 1.5-4.1 mm and 1.8-5.3 mm, respectively. The margin required to account for the setup error introduced using bony anatomy as a surrogate for the target, instead of localizing the target itself, was 0.5-1.4 cm. Conclusion: Using daily CBCT soft-tissue registration to localize the tumor in conventionally fractionated radiotherapy reduced the required setup margin by up to approximately 1.5 cm compared with both no image guidance and image guidance using bony anatomy as a surrogate for the target.« less