Fixed Exit Monochromator with fixed Rotation Axis
Journal Article
·
· AIP Conference Proceedings
- NSLS, Brookhaven National Laboratory, Upton, NY, 11973 (United States)
A new simple design for a fixed-exit monochromator has been developed. The set-up uses a linear slide to couple the rotation of the crystals to a translation of the second one to compensate for the 2hcos{theta} dependence of the beam-offset in a double crystal monochromator. This set-up requires just one motor for the rotation of the monochromator, and three piezo-actuators to tune the second crystal.The monochromator has been tested for Bragg-angles between 7 deg. and 70 deg.
- OSTI ID:
- 20653030
- Journal Information:
- AIP Conference Proceedings, Vol. 705, Issue 1; Conference: 8. international conference on synchrotron radiation instrumentation, San Francisco, CA (United States), 25-29 Aug 2003; Other Information: DOI: 10.1063/1.1757878; (c) 2004 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Compact Fixed-exit UHV DCM for XAFS
Variable exit beam height double-crystal monochromator for high-power insertion devices
Bent Crystal Monochromator with Constant Crystal Center Position and 2-theta Arm for a Dispersive Beamline
Journal Article
·
Fri Jan 19 00:00:00 EST 2007
· AIP Conference Proceedings
·
OSTI ID:20653030
+5 more
Variable exit beam height double-crystal monochromator for high-power insertion devices
Journal Article
·
Thu Feb 01 00:00:00 EST 1996
· Review of Scientific Instruments
·
OSTI ID:20653030
Bent Crystal Monochromator with Constant Crystal Center Position and 2-theta Arm for a Dispersive Beamline
Journal Article
·
Wed May 12 00:00:00 EDT 2004
· AIP Conference Proceedings
·
OSTI ID:20653030