skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Time dependent 14 MeV neutrons measurement using a polycrystalline chemical vapor deposited diamond detector at the JET tokamak

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1834691· OSTI ID:20644060
; ; ; ; ; ; ; ; ; ;  [1]
  1. Associazione EURATOM-ENEA sulla Fusione, ENEA C.R. Frascati, C.P. 65, I-00044 Frascati, Rome (Italy)

A polycrystalline chemical vapor deposited (CVD) diamond detector was installed on a JET tokamak in order to monitor the time dependent 14 MeV neutron emission produced by D-T plasma pulses during the Trace Tritium Experiment (TTE) performed in October 2003. This was the first tentative ever attempted to use a CVD diamond detector as neutron monitor in a tokamak environment. Despite its small active volume, the detector was able to detect the 14 MeV neutron emission (>1.0x10{sup 15} n/shot) with good reliability and stability during the experimental campaign that lasted five weeks. The comparison with standard silicon detectors presently used at JET as 14 MeV neutron monitors is reported, showing excellent correlation between the measurements. The results prove that CVD diamond detectors can be reliably used in a tokamak environment and therefore confirm the potential of this technology for next step machines like ITER.

OSTI ID:
20644060
Journal Information:
Review of Scientific Instruments, Vol. 76, Issue 1; Other Information: DOI: 10.1063/1.1834691; (c) 2005 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); JET EFDA Contributors; ISSN 0034-6748
Country of Publication:
United States
Language:
English