Towards increasing the performance of FTICR-MS with signal detection at frequency multiples: signal theory and numerical study

Journal Article · · International Journal of Mass Spectrometry

In Fourier transform mass spectrometry, detection of the signal at the cyclotron frequency multiples using multi-electrode ion traps, called nX-ICR cells, yields high resolution mass spectra at proportionally increased acquisition speed, and facilitates applications in omics and complex mixture analysis. In this approach, the mass-analyzer performance depends crucially on the trajectory of the ion motion during detection. For example, the ion orbital radius is one of the main factors affecting detection sensitivity nonlinearly, whereas high orbital radius during detection could significantly reduce the time of the ion coherent motion deteriorating both resolution and sensitivity. To address this issue and ensure optimal performance, a detailed description of the signal induced in the detection circuit of the nX-ICR cell for different ion motion trajectories and cell parameters is needed. In this study, we consider a simplified analytical model for image current induced on the electrodes of a model nX-ICR cell. The proposed formalism establishes the relationships between the cell parameters (e.g., detection circuit input resistance, temperature, frequency bandwidth, number of electrodes), and ion trajectory, and parameters of the acquired signal. These relationships reveal the key factors affecting the sensitivity of nX-ICR cell and intensities of undesirable peaks at the frequency multiples in the mass spectra.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
1819926
Report Number(s):
PNNL-SA-165728
Journal Information:
International Journal of Mass Spectrometry, Vol. 469
Country of Publication:
United States
Language:
English

References (51)

Accurate Mass Measurements in Proteomics journal August 2007
Precision proteomics: The case for high resolution and high mass accuracy journal September 2008
Petroleomics: advanced molecular probe for petroleum heavy ends journal March 2011
Evaluation of High-Field Orbitrap Fourier Transform Mass Spectrometer for Petroleomics journal June 2013
High Resolution Mass Spectrometry journal January 2012
High-Performance Mass Spectrometry: Fourier Transform Ion Cyclotron Resonance at 14.5 Tesla journal June 2008
Fourier Transform Mass Spectrometry journal July 2011
Applications of Fourier Transform Ion Cyclotron Resonance (FT-ICR) and Orbitrap Based High Resolution Mass Spectrometry in Metabolomics and Lipidomics journal May 2016
Fourier transform ion cyclotron resonance detection: principles and experimental configurations journal April 2002
Twelve Million Resolving Power on 4.7 T Fourier Transform Ion Cyclotron Resonance Instrument with Dynamically Harmonized Cell—Observation of Fine Structure in Peptide Mass Spectra journal March 2014
21 Tesla Fourier Transform Ion Cyclotron Resonance Mass Spectrometer Greatly Expands Mass Spectrometry Toolbox journal October 2016
Proteoform: a single term describing protein complexity journal February 2013
Phase Correction for Collision Model Analysis and Enhanced Resolving Power of Fourier Transform Ion Cyclotron Resonance Mass Spectra journal January 1999
Absorption mode Fourier transform mass spectrometry with no baseline correction using a novel asymmetric apodization function: Absorption mode FTMS with no baseline correction journal April 2015
Data processing in Fourier transform ion cyclotron resonance mass spectrometry: FT-ICR DATA PROCESSING journal January 2014
Application of frequency multiple FT-ICR MS signal acquisition for improved proteome research journal July 2021
21 Tesla Fourier Transform Ion Cyclotron Resonance Mass Spectrometer: A National Resource for Ultrahigh Resolution Mass Analysis journal June 2015
Advantageous Apodization Functions for Absorption-Mode Fourier Transform Spectroscopy journal May 1989
Absorption-Mode: The Next Generation of Fourier Transform Mass Spectra journal March 2012
Enhanced Fourier transform for Orbitrap mass spectrometry journal August 2014
Filter Diagonalization Method-Based Mass Spectrometry for Molecular and Macromolecular Structure Analysis journal March 2012
Use of the filter diagonalization method in the study of space charge related frequency modulation in Fourier transform ion cyclotron resonance mass spectrometry journal June 2006
From High- to Super-resolution Mass Spectrometry journal March 2014
Resolution improvement by using harmonic detection in an ion cyclotron resonance mass spectrometer journal November 1987
Harmonic signal enhancement in ion cyclotron resonance mass spectrometry using multiple electrode detection journal July 1988
High Speed Intact Protein Characterization Using 4X Frequency Multiplication, Ion Trap Harmonization, and 21 Tesla FTICR-MS journal April 2018
Can Fourier transform mass spectral resolution be improved by detection at harmonic multiples of the fundamental ion cyclotron orbital frequency? journal June 1991
Long-term coherence of the cyclotron mode in a trapped ion cloud journal May 1994
High-Resolution Fourier Transform Ion Cyclotron Resonance Mass Spectrometry with Increased Throughput for Biomolecular Analysis journal September 2014
Dynamics of ion motion in an elongated cylindrical cell of an ICR spectrometer and the shape of the signal registered journal March 1985
Cyclotron orbital radius determination in Fourier transform ion cyclotron resonance mass spectrometry journal May 1992
Analysis of harmonics for an elongated FTMS cell with multiple electrode detection journal December 1996
Simulations of nw measurement using multiple detection electrodes in FTICR mass spectrometry journal September 2020
Signal modeling for ion cyclotron resonance journal November 1978
Parallel Detection of Fundamental and Sixth Harmonic Signals Using an ICR Cell with Dipole and Sixth Harmonic Detectors journal January 2020
Fourier transform ion cyclotron resonance cell with dynamic harmonization of the electric field in the whole volume by shaping of the excitation and detection electrode assembly journal December 2010
Fourier transform ion cyclotron resonance (FT ICR) mass spectrometry: Theory and simulations: FT ICR MS journal February 2014
A novel high-performance Fourier transform ion cyclotron resonance cell for improved biopolymer characterization journal January 2000
The ‘hybrid cell’: a new compensated infinity cell for larger radius ion excitation in Fourier transform ion cyclotron resonance mass spectrometry journal January 2008
An electrically compensated trap designed to eighth order for FT-ICR mass spectrometry journal September 2008
Trapping Ring Electrode Cell: A FTICR Mass Spectrometer Cell for Improved Signal-to-Noise and Resolving Power journal August 2008
Trapped-ion cell with improved dc potential harmonicity for FT-ICR MS journal April 2008
Initial Experimental Characterization of a New Ultra-High Resolution FTICR Cell with Dynamic Harmonization journal April 2011
Ion behavior in an electrically compensated ion cyclotron resonance trap journal March 2011
Fourier transform ion cyclotron resonance mass spectrometry: the teenage years journal February 1991
Ion cyclotron resonance signal-detection at multiples of the cyclotron frequency journal May 1990
Comprehensive theory of the Fourier transform ion cyclotron resonance signal for all ion trap geometries journal April 1991
SIMION modeling of ion image charge detection in Fourier transform ion cyclotron resonance mass spectrometry journal June 2009
Theory of quadrupole detection fourier transform-ion cyclotron resonance journal July 1991
Fast image–charge calculations for multi-particle simulations in FT-ICR analyzer cells of arbitrary geometry journal February 2015
Fourier transform ion cyclotron resonance mass spectrometry: A primer journal January 1998

Related Subjects