Title: Calibrationless rotating Lorentz-force flowmeters for low flow rate applications

Dataset ·
DOI:https://doi.org/10.11578/1562007· OSTI ID:1562007

A 'weighted magnetic bearing' has been developed to improve the performance of rotating Lorentz-force flowmeters (RLFFs). Experiments have shown that the new bearing reduces frictional losses within a double-sided, disc-style RLFF to negligible levels. Operating such an RLFF under 'frictionless' conditions provides two major benefits. First, the steady-state velocity of the RLFF magnets matches the average velocity of the flowing liquid at low flow rates. This enables an RLFF to make accurate volumetric flow measurements without any calibration or prior knowledge of the fluid properties. Second, due to minimized frictional losses, an RLFF is able to measure low flow rates that cannot be detected when conventional, high-friction bearings are used. This paper provides a brief background on RLFFs, gives a detailed description of weighted magnetic bearings, and compares experimental RLFF data to measurements taken with a commercially available flowmeter.

Research Organization:
Princeton Plasma Physics Lab. (PPPL), Princeton, NJ (United States)
Sponsoring Organization:
U. S. Department of Energy
DOE Contract Number:
AC02-09CH11466
OSTI ID:
1562007
Country of Publication:
United States
Language:
English

References (1)

Calibrationless rotating Lorentz-force flowmeters for low flow rate applications journal May 2018

Cited By (1)

Calibrationless rotating Lorentz-force flowmeters for low flow rate applications journal May 2018

Similar Records

Calibrationless rotating Lorentz-force flowmeters for low flow rate applications
Journal Article · Tue May 29 00:00:00 EDT 2018 · Measurement Science and Technology · OSTI ID:1562007

Experimental calibration procedures for rotating Lorentz-force flowmeters
Journal Article · Fri Jul 14 00:00:00 EDT 2017 · Measurement Science and Technology · OSTI ID:1562007

Experimental calibration procedures for rotating Lorentz-force flowmeters
Dataset · Sun Jan 01 00:00:00 EST 2017 · OSTI ID:1562007