Test and characterization of 400 Hamamatsu R5912-MOD photomultiplier tubes for the ICARUS T600 detector
ICARUS T600 will be operated as far detector of the Short Baseline Neutrino program at Fermilab (U.S.A.), which foresees three liquid argon time projection chambers along the Booster Neutrino Beam line to search for a LSND-like sterile neutrino signal. The detector employs 360 photomultiplier tubes, Hamamatsu model R5912-MOD, suitable for cryogenic applications. A total of 400 PMTs were procured from Hamamatsu and tested at room temperature to evaluate the performance of the devices and their compliance to detect the liquid argon scintillation light in the T600 detector. Furthermore 60 units were also characterized at cryogenic temperature, in liquid argon bath, to evaluate any parameter variation which could affect the scintillation light detection. All the tested PMTs were found to comply with the requirements of ICARUS T600 and a subset of 360 specimens was selected for the final installation in the detector.
- Research Organization:
- Fermi National Accelerator Laboratory (FNAL), Batavia, IL (United States)
- Sponsoring Organization:
- USDOE Office of Science (SC), High Energy Physics (HEP)
- Contributing Organization:
- ICARUS/NP01 Collaboration
- Grant/Contract Number:
- AC02-07CH11359
- OSTI ID:
- 1478057
- Report Number(s):
- arXiv:1807.08577; FERMILAB-PUB-18-390-ND; 1683520
- Journal Information:
- Journal of Instrumentation, Vol. 13, Issue 10; ISSN 1748-0221
- Publisher:
- Institute of Physics (IOP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Web of Science
Design, construction and tests of the ICARUS T600 detector
|
journal | July 2004 |
Underground operation of the ICARUS T600 LAr-TPC: first results
|
journal | July 2011 |
Characterization of ETL 9357FLA photomultiplier tubes for cryogenic temperature applications
|
journal | January 2006 |
Realization of a high vacuum evaporation system for wave-length shifter deposition on photo-detector windows
|
journal | January 2018 |
Comparison between large area photo-multiplier tubes at cryogenic temperature for neutrino and rare event physics experiments
|
journal | July 2015 |
Characterization of large area PMTs at cryogenic temperature for rare event physics experiments
|
journal | March 2014 |
Timing properties of Hamamatsu R5912-MOD photomultiplier tube for the ICARUS T600 light detection system
|
journal | December 2018 |
Observation of light emission from Hamamatsu R11410-20 photomultiplier tubes
|
journal | September 2015 |
An innovative technique for TPB deposition on convex window photomultiplier tubes
|
journal | December 2018 |
Atmospheric Neutrino Search in the ICARUS T600 Detector
|
journal | January 2019 |
The ICARUS Experiment
|
journal | January 2019 |
The ICARUS experiment
|
conference | December 2003 |
An innovative technique for TPB deposition on convex window photomultiplier tubes | preprint | January 2018 |
Similar Records
Time calibration and synchronization of the scintillation light detection system in ICARUS-T600
The electronic set-up for the scintillation light detection system of ICARUS-SBN at Fermilab