Title: Rapid sample classification using an open port sampling interface coupled with liquid introduction atmospheric pressure ionization mass spectrometry

Abstract

An “Open Access”-like mass spectrometric platform to fully utilize the simplicity of the manual open port sampling interface for rapid characterization of unprocessed samples by liquid introduction atmospheric pressure ionization mass spectrometry has been lacking. The in-house developed integrated software with a simple, small and relatively low-cost mass spectrometry system introduced here fills this void. Software was developed to operate the mass spectrometer, to collect and process mass spectrometric data files, to build a database and to classify samples using such a database. These tasks were accomplished via the vendorprovided software libraries. Sample classification based on spectral comparison utilized the spectral contrast angle method. As a result, using the developed software platform near real-time sample classification is exemplified using a series of commercially available blue ink rollerball pens and vegetable oils. In the case of the inks, full scan positive and negative ion ESI mass spectra were both used for database generation and sample classification. For the vegetable oils, full scan positive ion mode APCI mass spectra were recorded. The overall accuracy of the employed spectral contrast angle statistical model was 95.3% and 98% in case of the inks and oils, respectively, using leave-one-out cross-validation. In conclusion, this work illustratesmore » that an open port sampling interface/mass spectrometer combination, with appropriate instrument control and data processing software, is a viable direct liquid extraction sampling and analysis system suitable for the non-expert user and near real-time sample classification via database matching.« less

Authors:
ORCiD logo [1]; ORCiD logo [1]
  1. Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
Publication Date:
Research Org.:
Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
Sponsoring Org.:
USDOE
OSTI Identifier:
1399227
Grant/Contract Number:  
AC05-00OR22725
Resource Type:
Journal Article: Accepted Manuscript
Journal Name:
Rapid Communications in Mass Spectrometry
Additional Journal Information:
Journal Volume: 31; Journal Issue: 3; Journal ID: ISSN 0951-4198
Publisher:
Wiley
Country of Publication:
United States
Language:
English
Subject:
37 INORGANIC, ORGANIC, PHYSICAL, AND ANALYTICAL CHEMISTRY; 97 MATHEMATICS AND COMPUTING

Citation Formats

Van Berkel, Gary J., and Kertesz, Vilmos. Rapid sample classification using an open port sampling interface coupled with liquid introduction atmospheric pressure ionization mass spectrometry. United States: N. p., 2016. Web. doi:10.1002/rcm.7792.
Van Berkel, Gary J., & Kertesz, Vilmos. Rapid sample classification using an open port sampling interface coupled with liquid introduction atmospheric pressure ionization mass spectrometry. United States. doi:10.1002/rcm.7792.
Van Berkel, Gary J., and Kertesz, Vilmos. Tue . "Rapid sample classification using an open port sampling interface coupled with liquid introduction atmospheric pressure ionization mass spectrometry". United States. doi:10.1002/rcm.7792. https://www.osti.gov/servlets/purl/1399227.
@article{osti_1399227,
title = {Rapid sample classification using an open port sampling interface coupled with liquid introduction atmospheric pressure ionization mass spectrometry},
author = {Van Berkel, Gary J. and Kertesz, Vilmos},
abstractNote = {An “Open Access”-like mass spectrometric platform to fully utilize the simplicity of the manual open port sampling interface for rapid characterization of unprocessed samples by liquid introduction atmospheric pressure ionization mass spectrometry has been lacking. The in-house developed integrated software with a simple, small and relatively low-cost mass spectrometry system introduced here fills this void. Software was developed to operate the mass spectrometer, to collect and process mass spectrometric data files, to build a database and to classify samples using such a database. These tasks were accomplished via the vendorprovided software libraries. Sample classification based on spectral comparison utilized the spectral contrast angle method. As a result, using the developed software platform near real-time sample classification is exemplified using a series of commercially available blue ink rollerball pens and vegetable oils. In the case of the inks, full scan positive and negative ion ESI mass spectra were both used for database generation and sample classification. For the vegetable oils, full scan positive ion mode APCI mass spectra were recorded. The overall accuracy of the employed spectral contrast angle statistical model was 95.3% and 98% in case of the inks and oils, respectively, using leave-one-out cross-validation. In conclusion, this work illustrates that an open port sampling interface/mass spectrometer combination, with appropriate instrument control and data processing software, is a viable direct liquid extraction sampling and analysis system suitable for the non-expert user and near real-time sample classification via database matching.},
doi = {10.1002/rcm.7792},
journal = {Rapid Communications in Mass Spectrometry},
number = 3,
volume = 31,
place = {United States},
year = {Tue Nov 15 00:00:00 EST 2016},
month = {Tue Nov 15 00:00:00 EST 2016}
}

Journal Article:
Free Publicly Available Full Text
Publisher's Version of Record

Citation Metrics:
Cited by: 9 works
Citation information provided by
Web of Science

Save / Share:

Works referenced in this record:

Versatile New Ion Source for the Analysis of Materials in Open Air under Ambient Conditions
journal, April 2005

  • Cody, Robert B.; Laramée, James A.; Durst, H. Dupont
  • Analytical Chemistry, Vol. 77, Issue 8, p. 2297-2302
  • DOI: 10.1021/ac050162j