skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Real-time dosimetry for boron neutron-capture therapy

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.467899· OSTI ID:136878
; ; ;  [1]
  1. Pacific Northwest Lab., Richland, WA (United States)

Epithermal/thermal boron neutron-capture therapy (BNCT) as a promising treatment method for malignant tumors. Because the doses rates for medical therapeutic radiation are very close to the normal tissue tolerance, small errors in radiation delivery can result in harmful overdoses. A substantial need exists for a device that will monitor, in real time, the radiation dose being delivered to a patient. Pacific Northwest Laboratory (PNL) has developed a scintillating glass optical fiber that is sensitive to thermal neutrons. The small size of the fibers offers the possibility of in vivo dose monitoring at several points within the radiation field. The count rate of such detectors may approach 10 MHz because the lifetime of the cerium activator is fast. Fluxes typical of those in BNCT (i.e., 10{sup 9} n/cm{sup 2}/sec) may be with a single fiber because of this potentially high count rate and the small diameter of the fiber.

OSTI ID:
136878
Report Number(s):
CONF-941061-; ISSN 0018-9499; TRN: 96:000980
Journal Information:
IEEE Transactions on Nuclear Science, Vol. 42, Issue 4Pt1; Conference: Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference, Norfolk, VA (United States), 30 Oct - 5 Nov 1994; Other Information: PBD: Aug 1995
Country of Publication:
United States
Language:
English