Title: Characterization of a spectroscopic detector for application in x-ray computed tomography

Abstract

Recent advances in cadmium telluride (CdTe) energy-discriminating pixelated detectors have enabled the possibility of Multi-Spectral X-ray Computed Tomography (MSXCT) to incorporate spectroscopic information into CT. MultiX ME 100 V2 is a CdTe-based spectroscopic x-ray detector array capable of recording energies from 20 to 160 keV in 1.1 keV energy bin increments. Hardware and software have been designed to perform radiographic and computed tomography tasks with this spectroscopic detector. Energy calibration is examined using the end-point energy of a bremsstrahlung spectrum and radioisotope spectral lines. When measuring the spectrum from Am-241 across 500 detector elements, the standard deviation of the peak-location and FWHM measurements are ±0.4 and ±0.6 keV, respectively. As these values are within the energy bin size (1.1 keV), detector elements are consistent with each other. The count rate is characterized, using a nonparalyzable model with a dead time of 64 ± 5 ns. This is consistent with the manufacturer’s quoted per detector-element linear-deviation at 2 Mpps (million photons per sec) of 8.9% (typical) and 12% (max). When comparing measured and simulated spectra, a low-energy tail is visible in the measured data due to the spectral response of the detector. If no valid photon detections are expected in themore » low-energy tail, then a background subtraction may be applied to allow for a possible first-order correction. If photons are expected in the low-energy tail, a detailed model must be implemented. A radiograph of an aluminum step wedge with a maximum height of about 20 mm shows an underestimation of attenuation by about 10% at 60 keV. This error is due to partial energy deposition from higher-energy (> 60 keV) photons into a lower-energy (~60 keV) bin, reducing the apparent attenuation. A radiograph of a PTFE cylinder taken using a bremsstrahlung spectrum from an x-ray voltage of 100 kV filtered by 1.3 mm Cu is reconstructed using Abel inversion. As no counts are expected in the low energy tail, a first order background correction is applied to the spectrum. The measured linear attenuation coefficient (LAC) is within 10% of the expected value in the 60 to 100 keV range. Below 60 keV, low counts in the corrected spectrum and partial energy deposition from incident photons of energy greater than 60 keV into energy bins below 60 keV impact the LAC measurements. This report ends with a demonstration of the tomographic capability of the system. The quantitative understanding of the detector developed in this report will enable further study in evaluating the system for characterization of an object’s chemical make-up for industrial and security purposes.« less

Authors:
 [1];  [1];  [1];  [1];  [1];  [1]
  1. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Publication Date:
Research Org.:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Org.:
USDOE
OSTI Identifier:
1368028
Report Number(s):
LLNL-TR-730135
DOE Contract Number:
AC52-07NA27344
Resource Type:
Technical Report
Country of Publication:
United States
Language:
English
Subject:
36 MATERIALS SCIENCE; 62 RADIOLOGY AND NUCLEAR MEDICINE; 47 OTHER INSTRUMENTATION

Citation Formats

Dooraghi, A. A., Fix, B. J., Smith, J. A., Brown, W. D., Azevedo, S. G., and Martz, H. E. Characterization of a spectroscopic detector for application in x-ray computed tomography. United States: N. p., 2017. Web. doi:10.2172/1368028.
Dooraghi, A. A., Fix, B. J., Smith, J. A., Brown, W. D., Azevedo, S. G., & Martz, H. E. Characterization of a spectroscopic detector for application in x-ray computed tomography. United States. doi:10.2172/1368028.
Dooraghi, A. A., Fix, B. J., Smith, J. A., Brown, W. D., Azevedo, S. G., and Martz, H. E. Fri . "Characterization of a spectroscopic detector for application in x-ray computed tomography". United States. doi:10.2172/1368028. https://www.osti.gov/servlets/purl/1368028.
@article{osti_1368028,
title = {Characterization of a spectroscopic detector for application in x-ray computed tomography},
author = {Dooraghi, A. A. and Fix, B. J. and Smith, J. A. and Brown, W. D. and Azevedo, S. G. and Martz, H. E.},
abstractNote = {Recent advances in cadmium telluride (CdTe) energy-discriminating pixelated detectors have enabled the possibility of Multi-Spectral X-ray Computed Tomography (MSXCT) to incorporate spectroscopic information into CT. MultiX ME 100 V2 is a CdTe-based spectroscopic x-ray detector array capable of recording energies from 20 to 160 keV in 1.1 keV energy bin increments. Hardware and software have been designed to perform radiographic and computed tomography tasks with this spectroscopic detector. Energy calibration is examined using the end-point energy of a bremsstrahlung spectrum and radioisotope spectral lines. When measuring the spectrum from Am-241 across 500 detector elements, the standard deviation of the peak-location and FWHM measurements are ±0.4 and ±0.6 keV, respectively. As these values are within the energy bin size (1.1 keV), detector elements are consistent with each other. The count rate is characterized, using a nonparalyzable model with a dead time of 64 ± 5 ns. This is consistent with the manufacturer’s quoted per detector-element linear-deviation at 2 Mpps (million photons per sec) of 8.9% (typical) and 12% (max). When comparing measured and simulated spectra, a low-energy tail is visible in the measured data due to the spectral response of the detector. If no valid photon detections are expected in the low-energy tail, then a background subtraction may be applied to allow for a possible first-order correction. If photons are expected in the low-energy tail, a detailed model must be implemented. A radiograph of an aluminum step wedge with a maximum height of about 20 mm shows an underestimation of attenuation by about 10% at 60 keV. This error is due to partial energy deposition from higher-energy (> 60 keV) photons into a lower-energy (~60 keV) bin, reducing the apparent attenuation. A radiograph of a PTFE cylinder taken using a bremsstrahlung spectrum from an x-ray voltage of 100 kV filtered by 1.3 mm Cu is reconstructed using Abel inversion. As no counts are expected in the low energy tail, a first order background correction is applied to the spectrum. The measured linear attenuation coefficient (LAC) is within 10% of the expected value in the 60 to 100 keV range. Below 60 keV, low counts in the corrected spectrum and partial energy deposition from incident photons of energy greater than 60 keV into energy bins below 60 keV impact the LAC measurements. This report ends with a demonstration of the tomographic capability of the system. The quantitative understanding of the detector developed in this report will enable further study in evaluating the system for characterization of an object’s chemical make-up for industrial and security purposes.},
doi = {10.2172/1368028},
journal = {},
number = ,
volume = ,
place = {United States},
year = {Fri Apr 21 00:00:00 EDT 2017},
month = {Fri Apr 21 00:00:00 EDT 2017}
}

Technical Report:

Save / Share:
  • It is proposed to build a Wolter X-ray Microscope Computed Tomography System in order to characterize objects to sub-micrometer resolution. Wolter Optics Systems use hyperbolic, elliptical, and/or parabolic mirrors to reflect x-rays in order to focus or magnify an image. Wolter Optics have been used as telescopes and as microscopes. As microscopes they have been used for a number of purposes such as measuring emission x-rays and x-ray fluoresce of thin biological samples. Standard Computed Tomography (CT) Systems use 2D radiographic images, from a series of rotational angles, acquired by passing x-rays through an object to reconstruct a 3D imagemore » of the object. The x-ray paths in a Wolter X-ray Microscope will be considerably different than those of a standard CT system. There is little information about the 2D radiographic images that can be expected from such a system. There are questions about the quality, resolution and focusing range of an image created with such a system. It is not known whether characterization information can be obtained from these images and whether these 2D images can be reconstructed to 3D images of the object. A code has been developed to model the 2D radiographic image created by an object in a Wolter X-ray Microscope. This code simply follows the x-ray through the object and optics. There is no modeling at this point of other effects, such as scattering, reflection losses etc. Any object, of appropriate size, can be used in the model code. A series of simulations using a number of different objects was run to study the effects of the optics. The next step will be to use this model to reconstruct an object from the simulated data. Funding for the project ended before this goal could be accomplished. The following documentation includes: (1) background information on current X-ray imaging systems, (2) background on Wolter Optics, (3) description of the Wolter System being used, (4) purpose, limitations and development of the modeling code, (5) description of the modeling code, (6) the results of a number of preliminary imaging simulations, and (7) recommendations for future Wolter designs and for further modeling studies.« less
  • The technology being proposed by LLNL is an Active and Passive Computed Tomography (A P CT) Drum Scanner for contact-handled (CH) wastes. It combines the advantages offered by two well-developed nondestructive assay technologies: gamma-ray spectrometry and computed tomography (CT). Coupled together, these two technologies offer to nondestructively and quantitatively characterize mixed- wastes forms. Gamma-ray spectroscopy uses one or more external radiation detectors to passively and nondestructively measure the energy spectrum emitted from a closed container. From the resulting spectrum one can identify most radioactivities detected, be they transuranic isotopes, mixed-fission products, activation products or environmental radioactivities. Spectral libraries exist atmore » LLNL for all four. Active (A) or transmission CT is a well-developed, nondestructive medical and industrial technique that uses an external-radiation beam to map regions of varying attenuation within a container. Passive (P) or emission CT is a technique mainly developed for medical application, e.g., single-photon emission CT. Nondestructive industrial uses of PCT are under development and just coming into use. This report discuses work on the A P CT Drum Scanner at LLNL.« less
  • This handbook presents typical primary x-ray spectra, as well as spectra which were obtained after transmission through water and lucite phantoms. Both types were obtained under good geometry conditions. These data were accumulated from four x-ray tubes under conditions that should simulate spectra encountered with current CT usage. Descriptive information is presented that should be helpful to the x-ray spectroscopist and that should aid the user in selecting the spectra that are most applicable to his condition. A discussion of the numerical data included in this handbook is presented. Graphic displays of incident exposure rates (mR/mAs at 1 meter) formore » some typical x-ray beams are presented and the data format used in this handbook is described. Appendix B contains numerical tabulations of the CT spectra.« less
  • A new technique using multiple flash x-ray sources for the determination of cross-sectional density profiles of rapidly translating or dynamically deforming objects is presented. The method, and extension of the concepts of medial tomography, allows stop-action survey of events with time constants at the low end of the microsecond range, making it suitable for problems of ballistic research and materials proof testing. This report discusses the design, the components, the capabilities, as well as the system studies which led up to the configuration which is under construction at BRL.
  • This report characterizes the current status of x-ray computed tomography (CT) as applied to the nondestructive evaluation of ceramic materials. The principal advantages of x-ray CT scanning are two. The first is the capability to produce high-contrast, geometrically accurate digital pictures (images) of slices through an object with a reasonably high spatial resolution. The second is that these slice images are not influenced to any significant degree by overlap or by structures outside of the plane of section (the tomographically imaged plane). Conventional radiography offers neither of these advantages. 34 refs., 42 figs., 4 tabs.