I reviewed all the hoods used for AS&I work in buildings 253, 254, and 255. All have had the minimum annual surveys conducted and some have had semi-annual reviews. The operating parameter labels are posted consistent to the right or above right (with the only exception the perchloric acid hood in 1734- on the left) which facilitates users knowing where to check for information. All hoods with sash doors (except the SE one in 1734 have a note on the Operating Parameters label to "keep sash doors closed". Again, this was checking consistency. No surveys were missed.