Systematic errors of long thin coaxial cables for the connection of four-terminal-pair devices
Conference
·
OSTI ID:125713
- Physikalisch-Technische Bundesanstalt, Braunschweig (Germany)
Four-terminal-pair devices which must be operated at liquid helium temperatures can only be connected to the measurement setup with relatively long, thin cables. This unfortunately leads to a large increase of the errors introduced by these cables. Some considerations are presented which allow these errors to be reduced as is necessary for precision measurements.
- OSTI ID:
- 125713
- Report Number(s):
- CONF-940603-; TRN: 95:005769-0080
- Resource Relation:
- Conference: CPEM `94: conference on precision electromagnetic measurements, Boulder, CO (United States), 27 Jun - 1 Jul 1994; Other Information: PBD: 1994; Related Information: Is Part Of 1994 Conference on precision electromagnetic measurements digest; DeWeese, E.; Bennett, G. [eds.]; PB: 577 p.
- Country of Publication:
- United States
- Language:
- English