Title: Operating experience review of an INL gas monitoring system

Abstract

This article describes the operations of several types of gas monitors in use at the Idaho National Laboratory (INL) High Temperature Electrolysis Experiment (HTE) laboratory. The gases monitored in the lab room are hydrogen, carbon monoxide, carbon dioxide, and oxygen. The operating time, calibration, and both actual and unwanted alarms are described. The calibration session time durations are described. In addition, some simple calculations are given to estimate the reliability of these monitors and the results are compared to operating experiences of other types of monitors.

Authors:
 [1];  [1];  [1]
  1. Idaho National Lab. (INL), Idaho Falls, ID (United States)
Publication Date:
Research Org.:
Idaho National Lab. (INL), Idaho Falls, ID (United States)
Sponsoring Org.:
USDOE
OSTI Identifier:
1184725
Report Number(s):
INL/JOU-14-31835
Journal ID: ISSN 1871-5532; PII: S1871553215000389
DOE Contract Number:  
AC07-05ID14517
Resource Type:
Journal Article
Resource Relation:
Journal Name: Journal of Chemical Health and Safety; Journal Volume: 42; Journal Issue: 8
Country of Publication:
United States
Language:
English
Subject:
47 OTHER INSTRUMENTATION; calibration; gas monitor; reliability

Citation Formats

Cadwallader, Lee C., DeWall, K. G., and Herring, J. S.. Operating experience review of an INL gas monitoring system. United States: N. p., 2015. Web. doi:10.1016/j.jchas.2015.02.002.
Cadwallader, Lee C., DeWall, K. G., & Herring, J. S.. Operating experience review of an INL gas monitoring system. United States. doi:10.1016/j.jchas.2015.02.002.
Cadwallader, Lee C., DeWall, K. G., and Herring, J. S.. Thu . "Operating experience review of an INL gas monitoring system". United States. doi:10.1016/j.jchas.2015.02.002.
@article{osti_1184725,
title = {Operating experience review of an INL gas monitoring system},
author = {Cadwallader, Lee C. and DeWall, K. G. and Herring, J. S.},
abstractNote = {This article describes the operations of several types of gas monitors in use at the Idaho National Laboratory (INL) High Temperature Electrolysis Experiment (HTE) laboratory. The gases monitored in the lab room are hydrogen, carbon monoxide, carbon dioxide, and oxygen. The operating time, calibration, and both actual and unwanted alarms are described. The calibration session time durations are described. In addition, some simple calculations are given to estimate the reliability of these monitors and the results are compared to operating experiences of other types of monitors.},
doi = {10.1016/j.jchas.2015.02.002},
journal = {Journal of Chemical Health and Safety},
number = 8,
volume = 42,
place = {United States},
year = {Thu Mar 12 00:00:00 EDT 2015},
month = {Thu Mar 12 00:00:00 EDT 2015}
}