Title: VALIDATION OF ANSI N42.34 AMERICAN NATIONAL STANDARD PERFORMANCE CRITERIA FOR HAND-HELD INSTRUMENTS FOR THE DETECTION AND IDENTIFICATION OF RADIONUCLIDES

Abstract

SRNL’s validation of ANSI N42.34-D6 for the Domestic Nuclear Detection Office (DNDO) was performed utilizing one hand-held instrument (or RID) – the FLIR identiFINDER 2. Each section of the standard was evaluated via a walk-through or test. NOTE: In Table 1, W = walk-through and T = test, as directed by the Domestic Nuclear Detection Office (DNDO). For a walk-through, the experiment was either setup or reviewed for setup; for a test, the N42.34-D6 procedures were followed with some exceptions and comments noted. SRNL is not fully able to evaluate a RID against Sections 7 (Environmental), 8 (Electromagnetic), and 9 (Mechanical) of N42.34, so those portions of this validation were done in collaboration with Qualtest, Inc. in Orlando, Florida. The walk-throughs and tests of Sections 7, 8, and 9 were performed in Qualtest, Inc. facilities with SRNL providing radiological sources as necessary. Where applicable, assessment results and findings of the walk-throughs and tests were recorded on datasheets and a validation summary is provided. A general comment pertained to test requirements found in another standard and referenced in N42.34-D6. For example, step 1 of the test method in section 8.1.2 states “RF test set up information can be found in IECmore » 61000-4-3.” It is recommended that any information from other standards necessary for conducting the tests within N42.34 should be posted in N42.34 for simplicity and to prevent the user from having to peruse other documents. Another general comment, as noted by Qualtest, is that a tolerance reference is not listed for each test in sections 7-9. Overall, the N42.34-D6 was proven to be practicable, but areas for improvement and recommendations were identified for consideration prior to final ballot submittal.« less

Authors:
Publication Date:
Research Org.:
SRS
Sponsoring Org.:
USDOE
OSTI Identifier:
1154716
Report Number(s):
SRNL-STI-2014-00360
DOE Contract Number:
DE-AC09-08SR22470
Resource Type:
Technical Report
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY

Citation Formats

Lorier, T. VALIDATION OF ANSI N42.34 AMERICAN NATIONAL STANDARD PERFORMANCE CRITERIA FOR HAND-HELD INSTRUMENTS FOR THE DETECTION AND IDENTIFICATION OF RADIONUCLIDES. United States: N. p., 2014. Web. doi:10.2172/1154716.
Lorier, T. VALIDATION OF ANSI N42.34 AMERICAN NATIONAL STANDARD PERFORMANCE CRITERIA FOR HAND-HELD INSTRUMENTS FOR THE DETECTION AND IDENTIFICATION OF RADIONUCLIDES. United States. doi:10.2172/1154716.
Lorier, T. Wed . "VALIDATION OF ANSI N42.34 AMERICAN NATIONAL STANDARD PERFORMANCE CRITERIA FOR HAND-HELD INSTRUMENTS FOR THE DETECTION AND IDENTIFICATION OF RADIONUCLIDES". United States. doi:10.2172/1154716. https://www.osti.gov/servlets/purl/1154716.
@article{osti_1154716,
title = {VALIDATION OF ANSI N42.34 AMERICAN NATIONAL STANDARD PERFORMANCE CRITERIA FOR HAND-HELD INSTRUMENTS FOR THE DETECTION AND IDENTIFICATION OF RADIONUCLIDES},
author = {Lorier, T.},
abstractNote = {SRNL’s validation of ANSI N42.34-D6 for the Domestic Nuclear Detection Office (DNDO) was performed utilizing one hand-held instrument (or RID) – the FLIR identiFINDER 2. Each section of the standard was evaluated via a walk-through or test. NOTE: In Table 1, W = walk-through and T = test, as directed by the Domestic Nuclear Detection Office (DNDO). For a walk-through, the experiment was either setup or reviewed for setup; for a test, the N42.34-D6 procedures were followed with some exceptions and comments noted. SRNL is not fully able to evaluate a RID against Sections 7 (Environmental), 8 (Electromagnetic), and 9 (Mechanical) of N42.34, so those portions of this validation were done in collaboration with Qualtest, Inc. in Orlando, Florida. The walk-throughs and tests of Sections 7, 8, and 9 were performed in Qualtest, Inc. facilities with SRNL providing radiological sources as necessary. Where applicable, assessment results and findings of the walk-throughs and tests were recorded on datasheets and a validation summary is provided. A general comment pertained to test requirements found in another standard and referenced in N42.34-D6. For example, step 1 of the test method in section 8.1.2 states “RF test set up information can be found in IEC 61000-4-3.” It is recommended that any information from other standards necessary for conducting the tests within N42.34 should be posted in N42.34 for simplicity and to prevent the user from having to peruse other documents. Another general comment, as noted by Qualtest, is that a tolerance reference is not listed for each test in sections 7-9. Overall, the N42.34-D6 was proven to be practicable, but areas for improvement and recommendations were identified for consideration prior to final ballot submittal.},
doi = {10.2172/1154716},
journal = {},
number = ,
volume = ,
place = {United States},
year = {Wed Sep 03 00:00:00 EDT 2014},
month = {Wed Sep 03 00:00:00 EDT 2014}
}

Technical Report:

Save / Share:
  • An evaluation of the neutron dosimeter used by the Centralized External Dosimetry System of Martin Marietta Energy Systems, Inc., was performed, and the dosimeter was shown to meet the requirements of the American National Standard for Personnel Neutron Dosimeters, ANSI N319-1976. This report details the requirements of the Standard, describes the tests performed, and evaluates the results of testing. To demonstrate compliance with the Standard, dosimeters were irradiated with a {sup 252}Cf source while mounted on a standard phantom. Dose was measured using the routine methodology employed by the Centralized External Dosimetry System for neutron dosimetry. The ability to accuratelymore » measure neutron dose was compared to specific performance criteria from the Standard. This includes testing the lower limit of detection, upper limit of detection, precision of results, and the capability to detect neutrons in a high gamma dose environment. In addition to neutron exposure, the dosimeters were required to be exposed to environmental factors including temperature extremes, high humidity, exposure to room light, and a drop to a hard surface. Only after exposure to these conditions were the dosimeters read, with results compared to the requirements of the Standard. Normal use factors of routine neutron dosimetry influencing the accuracy, sensitivity, or precision of the dosimetry system were also evaluated to measure their impact on dosimeter response.« less
  • American National Standards Institute Draft N13.29 describes performance tests for environmental radiation dosimetry providers. If approved it would be the first step toward applying the types of performance testing now required in personnel dosimetry to environmental radiation monitoring. The objective of this study was to pilot test the draft standard, before it undergoes final balloting, on a small group of dosimetry providers that were selected to provide a mix of facility types, thermoluminescent dosimeter designs and monitoring program applications. The first phase of the pilot test involved exposing dosimeters to laboratory photon, beta, and x-ray sources at routine and accidentmore » dose levels. In the second phase, dosimeters were subjected to ninety days of simulated environmental conditions in an environmental chamber that cycled through extremes of temperature and humidity. Two out of seven participants passed all categories of the laboratory testing phase, and all seven passed the environmental test phase. While some relatively minor deficiencies were uncovered in the course of the pilot test, the results show that draft N13.29 describes useful tests that could be appropriate for environmental dosimetry providers. An appendix to this report contains recommendations that should be addressed by the N13.29 working group before draft N13.29 is submitted for balloting.« less
  • In order to establish standards of bioassay performance upon which a uniform national program of performance testing might be based, the Health Physics Society Standards Committee (HPSSC) formed Working Group 2.5 to prepare the draft American National Standards Institute, Inc. (ANSI) Standard N13.30, Performance Criteria for Radiobioassay. Because the US Nuclear Regulatory Commission (NRC) is considering whether to require that all bioassay service laboratories meet the criteria of the draft Standard, the NRC staff requested that the Pacific Northwest Laboratory (PNL) estimate the costs that may be incurred in implementing the draft Standard. Two types of laboratories were involved inmore » the cost study: service laboratories and a performance testing laboratory. Cost estimates based on responses to questionnaires sent to seven facilities performing radiobioassays varied in relation to the extent of each facility's radiobioassay program and to their perception of their readiness for accreditation. Implementation of accreditation was estimated to cost each facility between a few thousand dollars to over one-quarter million dollars depending on the facility type. Likewise, annual costs ranged from a few hundred dollars to $170,000. For all but one facility, annual costs were estimated to be less than $25,000. In addition to these costs, startup cost for the testing laboratory were estimated to be approximately $322,000. 5 refs., 22 tabs.« less
  • This standard provides design criteria for systems and equipment necessary for the performance of those nuclear safety functions that are required to achieve and maintain a safe shutdown of the reactor to cold shutdown conditions from a hot standby or post accident condition. The following nuclear safety functions are addressed in this standard: reactivity control; reactor coolant system heat removal; reactor coolant system integrity; pressure control; and inventory control. The standard provides criteria for the design of systems used to accomplish these functions.
  • This standard provides the design, performance, testing, maintenance, and instrumentation and control requirements for the containment spray system for boiling water and pressurized water reactor stationary electric generating plants. The containment spray system, consisting of a spray subsystem and an additive subsystem, serves to mitigate the consequences of a loss of coolant accident, a PWR feedwater line or a PWR steam line break, by injecting a water spray into the containment atmosphere. The containment spray system may be used to aid safety by performing one or more of the following functions: (1) containment post-accident pressure suppression; (2) containment post-accident heatmore » removal; (3) containment atmosphere post-accident fission product removal; (4) post-accident mixing of containment atmosphere; (5) post-accident containment sump chemistry control.« less