Effects of vertical girder realignment in the Argonne APS storage ring.
The effects of vertical girder misalignments on the vertical orbit of the Advanced Photon Source (APS) storage ring are studied. Partial sector-realignment is prioritized in terms of the closed-orbit distortions due to misalignments of the corresponding girders in the sectors. A virtual girder-displacement (VGD) method is developed that allows the effects of a girder realignment to be tested prior to physically moving the girder. The method can also be used to anticipate the corrector strengths needed to restore the beam orbit after a realignment. Simulation results are compared to experimental results and found to reproduce the latter quite closely. Predicted corrector strengths are also found to be close to the actual local corrector strengths after a proof-of-principle two-sector realignment was performed.