skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: AN EVALUATION OF AN ELECTROCHEMICAL CELL AS AN OXYGEN GAUGE OVER A WIDE RANGE OF OXYGEN PRESSURES

Technical Report ·
DOI:https://doi.org/10.2172/12512903· OSTI ID:12512903

A galvanic cell using a commercially made calcium stabilized zirconia tube was used to measure the PO2 of analyzed and premixed gas mixtures of Ar-02 and CO-CO2 over a wide range of oxygen pressures near IOOO°C. The accuracy of the cell was evaluated by comparing these results with the reported analyses of these mixtures. A model based on thermal diffusion was used to explain the variation of the emf with flow at low flow rates. Errors due to a temperature difference across the electrolyte were also evaluated. When properly employed this cell has been shown to be a sensitive research device for the accurate measurement of the PO2 over a wide range of oxygen pressures.

Research Organization:
Metallurgy and Materials Science, College of Engineering, Marquette University, Milwaukee, WI
Sponsoring Organization:
US Atomic Energy Commission (AEC)
OSTI ID:
12512903
Report Number(s):
COO-1441-9
Country of Publication:
United States
Language:
English

Similar Records

A Study of the Defect Structure of Nonstoichiometric CeO2
Technical Report · Mon Sep 01 00:00:00 EDT 1969 · OSTI ID:12512903

Kinetics of Heterogeneous Reaction of CaCO3 Particles with Gaseous HNO3 Over a Wide Range of Humidity
Journal Article · Thu Feb 21 00:00:00 EST 2008 · Journal of Physical Chemistry A, 112(7):1561-1571 · OSTI ID:12512903

Development of a solid-electrolyte oxygen sensor for in situ oil shale retorting applications: final report
Technical Report · Wed Mar 01 00:00:00 EST 1978 · OSTI ID:12512903