CO (Carbon Monoxide Mixing Ratio System) Handbook
The main function of the CO instrument is to provide continuous accurate measurements of carbon monoxide mixing ratio at the ARM SGP Central Facility (CF) 60-meter tower (36.607 °N, 97.489 °W, 314 meters above sea level). The essential feature of the control and data acquisition system is to record signals from a Thermo Electron 48C and periodically calibrate out zero and span drifts in the instrument using the combination of a CO scrubber and two concentrations of span gas (100 and 300 ppb CO in air). The system was deployed on May 25, 2005.
- Research Organization:
- DOE Office of Science Atmospheric Radiation Measurement (ARM) Program (United States)
- Sponsoring Organization:
- USDOE EM Office of Science and Technology (EM-50)
- DOE Contract Number:
- DE-AC05-7601830
- OSTI ID:
- 1019542
- Report Number(s):
- DOE/SC-ARM/TR-072; TRN: US201120%%359
- Country of Publication:
- United States
- Language:
- English
Similar Records
Tower Temperature and Humidity Sensors (TWR) Handbook
NOAA Air Resources Laboratory Atmospheric Turbulence and Diffusion Division Contribution to LAFE Field Campaign Report
Tower Water-Vapor Mixing Ratio Value-Added Product Report
Technical Report
·
Mon Feb 01 00:00:00 EST 2010
·
OSTI ID:1019542
NOAA Air Resources Laboratory Atmospheric Turbulence and Diffusion Division Contribution to LAFE Field Campaign Report
Program Document
·
Thu Mar 22 00:00:00 EDT 2018
·
OSTI ID:1019542
+2 more
Tower Water-Vapor Mixing Ratio Value-Added Product Report
Technical Report
·
Tue Jun 01 00:00:00 EDT 2021
·
OSTI ID:1019542