skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: X-ray source brightness comparison: Rigaku rotating anode source vs. Kevex microfocus tube

Technical Report ·
DOI:https://doi.org/10.2172/974384· OSTI ID:974384

In 2007, we began to explore alternative x-ray sources for application to refraction-enhanced (phase contrast) x-ray radiography of cryogenic NIF ignition capsules containing frozen deuterium-tritium (D-T) ice layers. These radiographs are currently obtained using Kevex microfocus tubes as backlights, and for these sources the x-ray source size is approximately 5 {micro}m. As part of this exploration, we obtained refraction-enhanced radiographs of empty plastic capsules using the Janus laser facility at LLNL, demonstrating that even large ({approx} 100 {micro}m) sources can be utilized in refraction-enhanced radiography provided the source/sample distance is sufficiently large, and provided the final x-ray detector has sufficient spatial resolution. Essentially, in the current geometry, we rely on a small source to provide spatial resolution and on the source/sample distance to provide refraction contrast, but an equally useful alternative geometry is to use a large source and rely on fine detector spatial resolution to provide spatial resolution and on the sample/detector distance to provide refraction contrast.

Research Organization:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
974384
Report Number(s):
LLNL-TR-426090; TRN: US201009%%56
Country of Publication:
United States
Language:
English