Title: Stability of Philips XP2012 and XP2008 PMT's

Technical Report ·
DOI:https://doi.org/10.2172/913189· OSTI ID:913189

Results are presented for a 96-hour stability test of three XP2012 PMT's and one XP2008 PMT. These 1.5 inch diameter, 10-stage PMT's were used in the UA2 experiment and are candidates for some parts of the CDF plug upgrade calorimeter. The four PMT's tested appear to be just as stable as the Hamamatsu R4125's reported on in an earlier memo (TS-DET 93-024).

Research Organization:
Fermi National Accelerator Lab. (FNAL), Batavia, IL (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC02-07CH11359
OSTI ID:
913189
Report Number(s):
FERMILAB-MISC-1993-01; TRN: US200802%%367
Country of Publication:
United States
Language:
English