Title: Triple Coincidence Radioxenon Detector

Conference ·
OSTI ID:882591

The Automated Radioxenon Sampler/Analyzer (ARSA) built by Pacific Northwest National Laboratory (PNNL) is on e of the world’s most sensitive systems for monitoring the four radioxenon isotopes 133Xe, 133mXE, 131mXe and 135Xe. However, due to size, weight and power specifications appropriate to meet treaty-monitoring requirements; the ARSA is unsuitable for rapid deployment using modest transportation means. To transition this technology to a portable unit can be easily and rapidly deployed can be achieved by significant reductions in size, weight and power consumption if concentration were not required. As part of an exploratory effort to reduce both the size of the air sample and the gas processing requirement PNNL has developed an experimental nuclear detector to test and qualify the use of triple coincidence signatures (beta, conversion electron, x-ray) from two of the radioxenon isotopes (135Xe and 133Xe) as well as the more traditional beta-gamma coincidence signatures used by the ARSA system. The additional coincidence requirement allows for reduced passive shielding, and makes it possible for unambiguous detection of 133Xe and 153Xe in the presence of high 222Rn backgrounds. This paper will discuss the experimental setup and the results obtained for a 133Xe sample with and without 222Rn as an interference signature.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
882591
Report Number(s):
PNNL-SA-48661; NN2003000
Resource Relation:
Conference: Trends in Nuclear Explosion Monitoring, 26th Seismic Research Review: September 21-23, Orlando, Florida, LA-UR-04-5801:581-587
Country of Publication:
United States
Language:
English

Similar Records

Redesigned ß–gamma Radioxenon Detector
Journal Article · Tue Aug 21 00:00:00 EDT 2007 · Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:882591

Improved β-γ Coincidence Detector For Radioxenon Detection
Conference · Wed Aug 31 00:00:00 EDT 2005 · OSTI ID:882591

Discrimination Between Anthropogenic Sources of Atmospheric Radioxenon
Journal Article · Mon Dec 01 00:00:00 EST 2003 · Transactions of the American Nuclear Society, 87:89-90 · OSTI ID:882591