Detection and Quantification of Pu(III, IV, V, and VI) Using a1.0-meter Liquid Core Waveguide
Abstract
Detection and quantification of the aquo ions of Pu in 1 MHClO4 was carried out using a 1-meter liquid core waveguide (LCW) coupledto a fiber optic UV-Vis spectrometer. Detection limits of 7 x 10-7 M forPu(VI), 1.6 x 10-5 M for Pu(V), 5 x 10-6 M for Pu(IV) and 8 x 10-6 M forPu(III) were achieved. The limits of detection represent increases of 18to 33 times those achievable using a conventional 1-cm path length.Because of the much lower detection limits of the LCW, routineidentification of the oxidation states in dilute Pu solutions can bemade.
- Authors:
- Publication Date:
- Research Org.:
- Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
- Sponsoring Org.:
- USDOE Director. Office of Science. Office of CivilianRadioactive Waste Management
- OSTI Identifier:
- 882248
- Report Number(s):
- LBNL-54415
R&D Project: NHEAVY; BnR: KB0401024
- DOE Contract Number:
- DE-AC02-05CH11231
- Resource Type:
- Journal Article
- Journal Name:
- Radiochimica Acta
- Additional Journal Information:
- Journal Volume: 93; Related Information: Journal Publication Date: 2005
- Country of Publication:
- United States
- Language:
- English
- Subject:
- 38 RADIATION CHEMISTRY, RADIOCHEMISTRY, AND NUCLEAR CHEMISTRY; 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; plutonium absorption spectroscopy liquid corewaveguide
Citation Formats
Wilson, Richard E, Hu, Yung-Jin, and Nitsche, Heino. Detection and Quantification of Pu(III, IV, V, and VI) Using a1.0-meter Liquid Core Waveguide. United States: N. p., 2005.
Web. doi:10.1524/ract.93.4.203.64075.
Wilson, Richard E, Hu, Yung-Jin, & Nitsche, Heino. Detection and Quantification of Pu(III, IV, V, and VI) Using a1.0-meter Liquid Core Waveguide. United States. https://doi.org/10.1524/ract.93.4.203.64075
Wilson, Richard E, Hu, Yung-Jin, and Nitsche, Heino. 2005.
"Detection and Quantification of Pu(III, IV, V, and VI) Using a1.0-meter Liquid Core Waveguide". United States. https://doi.org/10.1524/ract.93.4.203.64075. https://www.osti.gov/servlets/purl/882248.
@article{osti_882248,
title = {Detection and Quantification of Pu(III, IV, V, and VI) Using a1.0-meter Liquid Core Waveguide},
author = {Wilson, Richard E and Hu, Yung-Jin and Nitsche, Heino},
abstractNote = {Detection and quantification of the aquo ions of Pu in 1 MHClO4 was carried out using a 1-meter liquid core waveguide (LCW) coupledto a fiber optic UV-Vis spectrometer. Detection limits of 7 x 10-7 M forPu(VI), 1.6 x 10-5 M for Pu(V), 5 x 10-6 M for Pu(IV) and 8 x 10-6 M forPu(III) were achieved. The limits of detection represent increases of 18to 33 times those achievable using a conventional 1-cm path length.Because of the much lower detection limits of the LCW, routineidentification of the oxidation states in dilute Pu solutions can bemade.},
doi = {10.1524/ract.93.4.203.64075},
url = {https://www.osti.gov/biblio/882248},
journal = {Radiochimica Acta},
number = ,
volume = 93,
place = {United States},
year = {Tue Feb 15 00:00:00 EST 2005},
month = {Tue Feb 15 00:00:00 EST 2005}
}
Other availability
Save to My Library
You must Sign In or Create an Account in order to save documents to your library.