skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: BGO Readout with Photodiodes as a Soft gamma-Ray Detector at -30C

Technical Report ·
DOI:https://doi.org/10.2172/839727· OSTI ID:839727

BGO is expected to be plausible devices for soft gamma-ray detectors, because of a high detection efficiency for soft gamma-rays. Here we report on the good performance of BGO readout with PIN-photodiode or avalanche photodiode as a soft gamma-ray detector. We confirmed that the signal output of BGO becomes comparable to that of GSO when it is readout with photodiodes due to better matching between emission wavelength of BGO and quantum efficiency of photodiode. The energy resolution of 6.2% and 3.4% for 662 keV and 1836 keV, respectively, gamma-rays at -30 C is obtained with the combination of the 5 x 5 x 5 mm{sup 3} cube BGO and the Hamamatsu avalanche photodiode (APD) S8664-55. In this combination, the lowest detectable energy is found to be {approx}10 keV. These performances are better than that obtained with Photomultiplier tube (PMT), and our results increase many possible applications of BGO readout with photodiodes as soft gamma-ray detectors.

Research Organization:
SLAC National Accelerator Lab., Menlo Park, CA (United States)
Sponsoring Organization:
USDOE Office of Science (SC)
DOE Contract Number:
AC02-76SF00515
OSTI ID:
839727
Report Number(s):
SLAC-PUB-10927; TRN: US0503312
Country of Publication:
United States
Language:
English