skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Neutron Measurements for Intensity Modulated Radiation Therapy

Technical Report ·
DOI:https://doi.org/10.2172/763769· OSTI ID:763769

The beam-on time for intensity modulated radiation therapy (IMRT) is increased significantly compared with conventional radiotherapy treatments. Further, the presence of beam modulation devices may potentially affect neutron production. Therefore, neutron measurements were performed for 15 MV photon beams on a Varian Clinac accelerator to determine the impact of IMRT on neutron dose equivalent to the patient.

Research Organization:
SLAC National Accelerator Lab., Menlo Park, CA (United States)
Sponsoring Organization:
USDOE Office of Energy Research (ER) (US)
DOE Contract Number:
AC03-76SF00515
OSTI ID:
763769
Report Number(s):
SLAC-PUB-8443; TRN: US0004805
Resource Relation:
Other Information: PBD: 21 Apr 2000
Country of Publication:
United States
Language:
English

Similar Records

Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy
Journal Article · Fri Jul 15 00:00:00 EDT 2005 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:763769

Investigation of conformal and intensity-modulated radiation therapy techniques to determine the absorbed fetal dose in pregnant patients with breast cancer
Journal Article · Fri Jul 01 00:00:00 EDT 2016 · Medical Dosimetry · OSTI ID:763769

The calculated risk of fatal secondary malignancies from intensity-modulated radiation therapy
Journal Article · Fri Jul 15 00:00:00 EDT 2005 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:763769