Title: Effect of Pb and air absorber thickness on [sup 137]Cs signal

Technical Report ·
DOI:https://doi.org/10.2172/6996378· OSTI ID:6996378

We present the results of a measurement to understand the effects of transverse misplacement of the [sup 137]Cs source tube inside the cast Pb plates of the SDC EMC calorimeter. The PMT current from a scintillator was measured as the [sup 137]Cs source was displaced by varying thicknesses of Pb, as well as varying thicknesses of air gap. At a nominal depth of 2 mm in Pb, we find a change in scintillator output of about 25%/mm, and about 10%/mm of air gap. The data are compared to a simple calculation. By taking the sum of the scintillators on top and below a source tube, we estimate a source tube displacement of [plus minus]450[mu] will cause an error in the PMT output by 1%.

Research Organization:
Argonne National Lab. (ANL), Argonne, IL (United States)
Sponsoring Organization:
USDOE; USDOE, Washington, DC (United States)
DOE Contract Number:
W-31109-ENG-38
OSTI ID:
6996378
Report Number(s):
ANL-HEP-TR-92-97; SDC-92-350; ON: DE93002783