Title: The stability of calibration standards for ICP/AES analysis: Six-month study

Technical Report ·
DOI:https://doi.org/10.2172/6923163· OSTI ID:6923163

The stability of instrument calibration standards for Inductively Coupled Plasma/Atomic Emission Spectrometric (ICP/AES) analysis was studied over a six-month period. Data were obtained as functions of analyte concentration, acid type, and acidity. The impact of acid concentration on signal-to-background ratios (S/B) was also assessed. The results show that analytes maintain their integrity over extended periods with appropriate inorganic acid preservatives. Thus, frequent standard preparations become unnecessary to obtain valid analytical data.

Research Organization:
Argonne National Lab., IL (United States). Analytical Chemistry Lab.
Sponsoring Organization:
USDOE; USDOE, Washington, DC (United States)
DOE Contract Number:
W-31109-ENG-38
OSTI ID:
6923163
Report Number(s):
ANL/ACL-92/3; ON: DE93003217
Country of Publication:
United States
Language:
English