Title: Design considerations for an x-ray microprobe

Conference ·
OSTI ID:6794322

The optical design of a fluorescent microprobe covering the x-ray region from 2 to 16 keV is considered for the NSLS x-ray ring. The limit on detectability is from total flux (photons/..mu..m/sup 2/) and several design choices are considered to match the optical system to the storage ring to maximize throughput. The tradeoffs in image quality and energy resolution of these designs have been considered and within these constraints two firm proposals are presented.

Research Organization:
Brookhaven National Lab., Upton, NY (USA)
DOE Contract Number:
AC02-76CH00016
OSTI ID:
6794322
Report Number(s):
BNL-31971; CONF-820839-4; ON: DE83000976
Resource Relation:
Conference: International conference on X-ray and VUV synchrotron radiation instrumentation, Hamburg, F.R. Germany, 9 Aug 1982; Other Information: Portions of document are illegible