Potential improvements in instrumentation for PET
Conference
·
OSTI ID:6380127
This paper discusses the potential for improved detectors in Positron Emission Tomography (PET) and explores the ultimate limits that might be achieved in the areas of spatial resolution, sensitivity, and maximum imaging rates. It is shown that if an ultra-fast, high efficiency scintillator and a thin, low-noise, position-sensitive photodetector were available, a multi-layer time-of-flight tomograph would be possible with a 10 cm axial field of view, a 3-dimensional spatial resolution of 2 mm fwhm, and >700,000 prompt unscattered coincidences per sec for 1 ..mu..Ci per cm/sup 3/ in a 20 cm diam cylinder of water. 39 refs., 3 figs., 3 tabs.
- Research Organization:
- Lawrence Berkeley Lab., CA (USA)
- DOE Contract Number:
- AC03-76SF00098
- OSTI ID:
- 6380127
- Report Number(s):
- LBL-20215; CONF-8409164-9; ON: DE86002893
- Resource Relation:
- Conference: NATO advanced study institute on physics and engineering of medical imaging, Maratea, Italy, 24 Sep 1984
- Country of Publication:
- United States
- Language:
- English
Similar Records
Monte Carlo calculations of PET coincidence timing: single and double-ended readout
HEADTOME III: A high quantitation and high resolution brain positron emission tomograph
Evaluation of cerium activated gadolinium orthosilicate (GSO) as a detector for PET
Journal Article
·
Tue Sep 08 00:00:00 EDT 2015
· Physics in Medicine and Biology
·
OSTI ID:6380127
HEADTOME III: A high quantitation and high resolution brain positron emission tomograph
Conference
·
Sun Jan 01 00:00:00 EST 1984
· J. Nucl. Med.; (United States)
·
OSTI ID:6380127
+3 more
Evaluation of cerium activated gadolinium orthosilicate (GSO) as a detector for PET
Conference
·
Sun Jan 01 00:00:00 EST 1984
· J. Nucl. Med.; (United States)
·
OSTI ID:6380127
+2 more
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
POSITION SENSITIVE DETECTORS
SENSITIVITY
POSITRON COMPUTED TOMOGRAPHY
SPATIAL RESOLUTION
SOLID SCINTILLATION DETECTORS
TIME-OF-FLIGHT METHOD
COMPUTERIZED TOMOGRAPHY
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
MEASURING INSTRUMENTS
RADIATION DETECTORS
RESOLUTION
SCINTILLATION COUNTERS
TOMOGRAPHY
550601* - Medicine- Unsealed Radionuclides in Diagnostics
POSITION SENSITIVE DETECTORS
SENSITIVITY
POSITRON COMPUTED TOMOGRAPHY
SPATIAL RESOLUTION
SOLID SCINTILLATION DETECTORS
TIME-OF-FLIGHT METHOD
COMPUTERIZED TOMOGRAPHY
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
MEASURING INSTRUMENTS
RADIATION DETECTORS
RESOLUTION
SCINTILLATION COUNTERS
TOMOGRAPHY
550601* - Medicine- Unsealed Radionuclides in Diagnostics