skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Potential improvements in instrumentation for PET

Conference ·
OSTI ID:6380127

This paper discusses the potential for improved detectors in Positron Emission Tomography (PET) and explores the ultimate limits that might be achieved in the areas of spatial resolution, sensitivity, and maximum imaging rates. It is shown that if an ultra-fast, high efficiency scintillator and a thin, low-noise, position-sensitive photodetector were available, a multi-layer time-of-flight tomograph would be possible with a 10 cm axial field of view, a 3-dimensional spatial resolution of 2 mm fwhm, and >700,000 prompt unscattered coincidences per sec for 1 ..mu..Ci per cm/sup 3/ in a 20 cm diam cylinder of water. 39 refs., 3 figs., 3 tabs.

Research Organization:
Lawrence Berkeley Lab., CA (USA)
DOE Contract Number:
AC03-76SF00098
OSTI ID:
6380127
Report Number(s):
LBL-20215; CONF-8409164-9; ON: DE86002893
Resource Relation:
Conference: NATO advanced study institute on physics and engineering of medical imaging, Maratea, Italy, 24 Sep 1984
Country of Publication:
United States
Language:
English