Title: An isotopic analysis system for plutonium samples enriched in sup 238 Pu

Conference ·
OSTI ID:6243658

We have designed and built a gamma-ray spectrometer system that measures the relative plutonium isotopic abundances of plutonium oxide enriched in {sup 238}Pu. The first system installed at Westinghouse Savannah River Company was tested and evaluated on plutonium oxide in stainless steel EP60/61 containers. {sup 238}Pu enrichments ranged from 20% to 85%. Results show that 200 grams of plutonium oxide in an EP60.61 container can be measured with {plus minus}0.3% precision and better than {plus minus}1.0% accuracy in the specific power using a counting time of 50 minutes. 3 refs., 2 figs.

Research Organization:
Lawrence Livermore National Lab., CA (United States)
Sponsoring Organization:
USDOE; USDOE, Washington, DC (United States)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
6243658
Report Number(s):
UCRL-JC-105887; CONF-910901-21; ON: DE92003067; TRN: 91-033240
Resource Relation:
Conference: 4. international conference on facility operations/safeguards interface, Albuquerque, NM (United States), 29 Sep - 4 Oct 1991
Country of Publication:
United States
Language:
English