Title: ICP-MS nebulizer performance for analysis of SRS high salt simulated radioactive waste tank solutions ({number_sign}3053)

Conference ·
OSTI ID:568053

High Level Radioactive Waste Tanks at the Savannah River Site are high in salt content. The cross-flow nebulizer provided the most stable signal for all salt matrices with the smallest signal loss/suppression due to this matrix. The DIN exhibited a serious lack of tolerance for TDS; possibly due to physical de-tuning of the nebulizer efficiency.

Research Organization:
Savannah River Site (SRS), Aiken, SC (United States)
Sponsoring Organization:
USDOE Office of Environmental Restoration and Waste Management, Washington, DC (United States)
DOE Contract Number:
AC09-96SR18500
OSTI ID:
568053
Report Number(s):
WSRC-MS-97-00790; CONF-980317-; ON: DE98050691; TRN: 98:002563
Resource Relation:
Conference: 2. ASTM symposium on applications of ICP-MS to radionuclide determinations, New Orleans, LA (United States), 2-7 Mar 1998; Other Information: PBD: [1997]