Nanosecond fluorescence spectroscopy
This article is a summary of a short course lecture given in conjunction with the 1984 Nuclear Science Symposium. Measuring systems for nanosecond fluorescence spectroscopy using single-photon counting techniques are presented. These involve systems based on relaxation-type spark gap light pulser and synchronously pumped mode-locked dye lasers. Furthermore, typical characteristics and optimization of operating conditions of the critical components responsible for the system time resolution are discussed. A short comparison of the most important deconvolution methods for numerical analysis of experimental data is given particularly with respect to the signal-to-noise ratio of the fluorescence signal. 22 refs., 8 figs.
- Research Organization:
- Lawrence Berkeley Lab., CA (USA)
- DOE Contract Number:
- AC03-76SF00098
- OSTI ID:
- 5557051
- Report Number(s):
- LBL-19291; CONF-851097-1; ON: DE85012234
- Resource Relation:
- Conference: IEEE symposium, San Francisco, CA, USA, 23 Oct 1985
- Country of Publication:
- United States
- Language:
- English
Similar Records
Balanced cross-rate model for saturated molecular fluorescence in flames using a nanosecond pulse length laser
Multi-channel lock-in amplifier assisted femtosecond time-resolved fluorescence non-collinear optical parametric amplification spectroscopy with efficient rejection of superfluorescence background
Related Subjects
FLUORESCENCE SPECTROSCOPY
PULSE TECHNIQUES
MOLECULAR STRUCTURE
MEASURING METHODS
PULSE ANALYZERS
RESOLUTION
REVIEWS
TIME DEPENDENCE
TIMING CIRCUITS
DOCUMENT TYPES
ELECTRONIC CIRCUITS
ELECTRONIC EQUIPMENT
EMISSION SPECTROSCOPY
EQUIPMENT
SPECTROSCOPY
440300* - Miscellaneous Instruments- (-1989)