Title: Automatic network analyzer procedures for 5045 klystron cavities

Technical Report ·
DOI:https://doi.org/10.2172/5275580· OSTI ID:5275580

This Note describes the results of using Automatic Network Analyzers in measuring SLAC 5045 klystron cavities. Two different analyzers were compared; the HP8753 and HP8510. Both analyzers have frequency synthesizer accuracy and stability to perform the measurement without the need for a frequency counter. The klystron has six cavities which can be put into three categories; input, gain and output. The input and output cavities require an external Q measurement (Qe) to determine coupling ({beta}) and center frequency (f{sub o}). The gain cavities require a resonant frequency measurement only.