A hybrid digital-analog long pulse integrator
A digital-analog integrator has been developed for use with inductive magnetic sensors in long-pulse tokamaks. Continuous compensation of input offsets is accomplished by alternating analog-to-digital convertor samples from the sensor and a dummy load, while an RC network provides passive integration between samples. Typically a sampling rate of 10 kHz is used. In operational tests on the DIII-D tokamak, digital and analog integration of tokamak data show good agreement. The output drift error during a 1200 s integration interval corresponds to a few percent of the anticipated signal for poloidal field probes in International Thermonuclear Experimental Reactor (ITER), and bench tests suggest that the error can be reduced further.
- Research Organization:
- General Atomics, San Diego, CA (United States)
- Sponsoring Organization:
- USDOE Office of Energy Research, Washington, DC (United States)
- DOE Contract Number:
- AC03-89ER51114
- OSTI ID:
- 448065
- Report Number(s):
- GA-A22338; CONF-960543-44; ON: DE97004335; TRN: 97:006552
- Resource Relation:
- Conference: 11. annual high-temperature plasma diagnostics conference, Monterey, CA (United States), 12-16 May 1996; Other Information: PBD: May 1996
- Country of Publication:
- United States