Title: The tomography beamline at the National Synchrotron Light Source

Conference ·
OSTI ID:434926

We compared the image contrast of a monochromatic CT, Multiple Energy Computed Tomography (MECT), and conventional CT scanner using phantoms. The experimental results indicate that monochromatic CT, with beam energy tuned just above the iodine K-edge, has about a 3 fold advantage in iodine contrast over conventional CT with a 120 kVp beam. Modeling using the same beams at a 3 rad dose and 3 mm slice height on an 18 cm diameter acrylic phantom, the simulations show a noise of 1.2 HU for MECT and 1.9 HU for CCT. Furthermore, despite the Cupping-effect corrections the bone contrast is lower in CCT and varies by 24 HU moving from the phantom`s center to the edge; this indicates an advantage for MECT in detecting and quantifying lesions differing from surrounding tissue by their mean atomic number.

Research Organization:
Brookhaven National Lab. (BNL), Upton, NY (United States)
Sponsoring Organization:
USDOE Office of Energy Research, Washington, DC (United States)
DOE Contract Number:
AC02-76CH00016
OSTI ID:
434926
Report Number(s):
BNL-63611; CONF-960955-2; ON: DE97000979
Resource Relation:
Conference: 5. international conference on applications of physics in medicine and biology, Trieste (Italy), 2-6 Sep 1996; Other Information: PBD: [1996]
Country of Publication:
United States
Language:
English