A simple, low-cost, versatile CCD spectrometer for plasma spectroscopy
- Univ. of Wisconsin, Madison, WI (United States). Dept. of Physics
- Sterling Scientific, Inc., Madison, WI (United States)
The authors have constructed a simple, low-cost CCD spectrometer capable of both high resolution ({Delta}{lambda} {le} 0.015 nm) and large bandpass (110 nm with {Delta}{lambda} {approximately}0.3 nm). These two modes of operation provide two broad areas of capability for plasma spectroscopy. The first major application is measurement of emission line broadening; the second is emission line surveys from the ultraviolet to the near infrared. Measurements have been made on a low-temperature plasma produced by a miniature electrostatic plasma source and the high-temperature plasma in the MST Reversed-Field Pinch. The spectrometer is a modified Jarrell-Ash 0.5 m Ebert-Fastie monochromator. Light is coupled into the entrance slit with a fused silica fiber optic bundle. The exposure time (2 ms minimum) is controlled by a fast electromechanical shutter. The exit plane detector is a compact and robust CCD detector developed for amateur astronomy by Santa Barbara Instrument Group. The CCD detector is controlled and read out by a Macintosh{reg_sign} computer. This spectrometer is sophisticated enough to serve well in a research laboratory, yet is simple and inexpensive enough to be affordable for instructional use.
- Research Organization:
- Wisconsin Univ., Madison, WI (United States). Plasma Physics Research
- Sponsoring Organization:
- USDOE, Washington, DC (United States)
- DOE Contract Number:
- FG02-85ER53198
- OSTI ID:
- 266729
- Report Number(s):
- DOE/ER/53198-273; CONF-960543-27; ON: DE96012450; TRN: 96:017026
- Resource Relation:
- Conference: 11. annual high-temperature plasma diagnostics conference, Monterey, CA (United States), 12-16 May 1996; Other Information: PBD: Jun 1996
- Country of Publication:
- United States
- Language:
- English