Title: Characterization of X-ray fields at the center for devices and radiological health

Conference ·
OSTI ID:250236
 [1]
  1. Center for Devices and Radiological Health, Rockville, MD (United States)

This talk summarizes the process undertaken by the Center for Devices and Radiological Health (CDRH) for establishing reference x-ray fields in its accredited calibration laboratory. The main considerations and their effects on the calibration parameters are discussed. The characterization of fields may be broken down into two parts: (1) the initial setup of the calibration beam spectra and (2) the ongoing measurements and controls which ensure consistency of the reference fields. The methods employed by CDRH for both these stages and underlying considerations are presented. Uncertainties associated with the various parameters are discussed. Finally, the laboratory`s performance, as evidenced by ongoing measurement quality assurance results, is reported.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
OSTI ID:
250236
Report Number(s):
PNL-10076; CONF-9303220-; ON: DE96008816; TRN: 96:002617-0028
Resource Relation:
Conference: Workshop on measurement quality assurance for ionizing radiation, Gaithersburg, MD (United States), 16-18 Mar 1993; Other Information: PBD: [1993]; Related Information: Is Part Of Workshop on measurement quality assurance for ionizing radiation: Proceedings; Heath, J.A.; Swinth, K.L. [comps.]; PB: 468 p.