Title: Isotope Exchange and Fractionation Corrections for Extraction of Tritiated Water in Silica Gel by Freeze-Drying Techniques

Technical Report ·
DOI:https://doi.org/10.2172/15013462· OSTI ID:15013462

A concentration correction curve was established for measuring the activity concentration of airborne tritiated water collected with dried silica gel and extracted by the LLNL Environmental Monitoring Radiological Laboratory freeze-dry technique. A tracer study using standard tritiated water with silica gel showed that the concentration of tritium in the extracted water is lower than that in the adsorbed water by a fraction proportional to the amount of adsorbed water. The observed decrease in tritium concentration in the extracted water can be accounted for by dilution due to isotopic exchange with both non-tritiated water and hydroxyl groups within the silica gel matrix. For the range of 8-35% adsorbed water, which is typical of samples collected in LLNL monitoring stations, the derived exchangeable water in the silica gel material under investigation was (5.12 {+-} 0.08)%. The contribution of the H{sub 2}O/HTO vapor pressure effect using published empirical data in the literature was also considered in calculating the degree of isotopic exchange.

Research Organization:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
15013462
Report Number(s):
UCRL-ID-148610; TRN: US200601%%458
Country of Publication:
United States
Language:
English