Establishment of Harrop, High-Temperature Viscometer
This report explains how the Harrop, High-Temperature Viscometer was installed, calibrated, and operated. This report includes assembly and alignment of the furnace, viscometer, and spindle, and explains the operation of the Brookfield Viscometer, the Harrop furnace, and the UDC furnace controller. Calibration data and the development of the spindle constant from NIST standard reference glasses is presented. A simple operational procedure is included.
- Research Organization:
- Savannah River Site (SRS), Aiken, SC (United States)
- Sponsoring Organization:
- US Department of Energy (US)
- DOE Contract Number:
- AC09-96SR18500
- OSTI ID:
- 14492
- Report Number(s):
- WSRC-RP-98-00737; TRN: US0111056
- Resource Relation:
- Other Information: PBD: 5 Nov 1999
- Country of Publication:
- United States
- Language:
- English