A comparative study of 129I content in environmental standard materials IAEA-375, NIST SRM 4354 and NIST SRM 4357 by Thermal Ionization Mass Spectrometry and Accelerator Mass Spectrometry
Iodine environmental measurements have consistently been backed up in the literature by standard materials like IAEA-375, Chernobyl Soil. There are not many other sources of a certified reference material for 129I content for mass spectrometry measurements. Some that have been found in the literature include NIST-4354 and NIST-4357. They are still available at the time of this writing. They don’t have certified content or isotopic values. There has been some work in the literature to show that iodine is present, but there hasn’t been enough to establish a consensus value. These materials have been analyzed at INL through two separate mass spectrometry techniques. They involve a combustion method of the starting material in oxygen, followed by TIMS analysis and a leaching preparation analyzed by accelerator mass spectrometry. Combustion/TIMS preparation of NIST SRM-4354 resulted in a 129I/127I ratio of 1.92 x 10-6 which agrees with AMS measurements which measured the 129I/127I ratio to be 1.93 x 10-6.
- Research Organization:
- Idaho National Lab. (INL), Idaho Falls, ID (United States)
- Sponsoring Organization:
- USDOE Office of Nuclear Energy (NE)
- DOE Contract Number:
- DE-AC07-05ID14517
- OSTI ID:
- 1367514
- Report Number(s):
- INL/CON-16-38652
- Resource Relation:
- Conference: 2016 Low-Level Radioactivity Measurement Techniques (LLRMT) Conference, Seattle, Washington, USA, September 26–30, 2016
- Country of Publication:
- United States
- Language:
- English
Similar Records
Determination of iodine in oyster tissue by isotope dilution laser resonance ionization mass spectrometry
Transition of Iodine Analysis to Accelerator Mass Spectrometry