Performance Evaluation of Spectroscopic Detectors for LEU Hold-up Measurements
- Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
The hold-up measurement of low-enriched uranium materials may require use of alternate detector types relative to the measurement of highly enriched uranium. This is in part due to the difference in process scale (i.e., the components are generally larger for low-enriched uranium systems), but also because the characteristic gamma-ray lines from 235U used for assay of highly enriched uranium will be present at a much reduced intensity (on a per gram of uranium basis) at lower enrichments. Researchers at Oak Ridge National Laboratory examined the performance of several standard detector types, e.g., NaI(Tl), LaBr3(Ce), and HPGe, to select a suitable candidate for measuring and quantifying low-enriched uranium hold-up in process pipes and equipment at the Portsmouth gaseous diffusion plant. Detector characteristics, such as energy resolution (full width at half maximum) and net peak count rates at gamma ray energies spanning a range of 60–1332 keV, were measured for the above-mentioned detector types using the same sources and in the same geometry. Uranium enrichment standards (Certified Reference Material no. 969 and Certified Reference Material no. 146) were measured using each of the detector candidates in the same geometry. The net count rates recorded by each detector at 186 keV and 1,001 keV were plotted as a function of enrichment (atom percentage). Background measurements were made in unshielded and shielded configurations under both ambient and elevated conditions of 238U activity. The highly enriched uranium hold-up measurement campaign at the Portsmouth plant was performed on process equipment that had been cleaned out. Therefore, in most cases, the thickness of the uranium deposits was less than the “infinite thickness” for the 186 keV gamma rays to be completely self-attenuated. Because of this, in addition to measuring the 186 keV gamma, the 1,001 keV gamma ray from 234mPa—a daughter of 238U in secular equilibrium with its parent—will also need to be measured. Based on the performance criteria of detection efficiency, energy resolution, peak-to-continuum ratios, minimum detectable limits, and the weight of the shielded probe, a shielded (0.5 in. thick lead shield) 2 × 2 in. NaI(Tl) detector is recommended for use. The recommended approach is to carry out analysis using data from both 186 keV and 1,001 keV gamma rays, and select a best result based on propagated uncertainty estimates. It is also highly recommended that a two-point gain stabilization scheme based on an 241Am seed embedded in the probe be implemented. Shielding configurations to reduce the impact of background interference on the measurement of 1,001 keV gamma-ray are discussed.
- Research Organization:
- Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States). Quantum Condensed Matter Division
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC05-00OR22725
- OSTI ID:
- 1335357
- Report Number(s):
- ORNL/TM-2016/649; TRN: US1700828
- Country of Publication:
- United States
- Language:
- English
Similar Records
MODEL AND ALGORITHM EVALUATION FOR THE HYBRID UF6 CONTAINER INSPECTION SYSTEM
Stationary and protable instruments for assay of HEU (highly enriched uranium) solids holdup
Related Subjects
SLIGHTLY ENRICHED URANIUM
GAMMA RADIATION
PORTSMOUTH GASEOUS DIFFUSION PLANT
KEV RANGE 100-1000
MEV RANGE 01-10
NAI DETECTORS
HIGH-PURITY GE DETECTORS
URANIUM 238
PROTACTINIUM 234
ENERGY RESOLUTION
PERFORMANCE TESTING
THICKNESS
LEAD
COUNTING RATES
ENRICHMENT
EVALUATION
GAIN
PEAKS
PIPES
PROBES
STANDARDS
EFFICIENCY
INTERFERENCE
SHIELDS
SEMICONDUCTOR DETECTORS
LANTHANUM BROMIDES
Low Enriched Uranium
Gamma Ray Holdup