skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: CHARACTERIZING SURFACE LAYERS IN NITINOL USING X-RAY PHOTOELECTRON SPECTROSCOPY

Journal Article · · Journal of Undergraduate Research
OSTI ID:1052084

Nitinol is a shape memory alloy whose properties allow for large reversible deformations and a return to its original geometry. This nickel-titanium (NiTi) alloy has become a material used widely in the biomedical fi eld as a stent to open up collapsed arteries. Both ambient and biological conditions cause surface oxidation in these devices which in turn change its biocompatibility. The thickness of oxidized layers can cause fractures in the material if too large and can allow for penetration if too thin. Depending on the type and abundance of the chemical species on or near the surface, highly toxic metal ions can leak into the body causing cell damage or even cell death. Thus, biocompatibility of such devices is crucial. By using highly surface sensitive x-ray photoelectron spectroscopy to probe the surface of these structures, it is possible to decipher both layer composition and layer thickness. Two samples, both of which were mechanically polished, were investigated. Of the two samples, one was then exposed to a phosphate buffered saline (PBS) solution to mimic the chemical properties of blood, while the other remained unexposed. Although both samples were found to have oxide layers of appropriate thickness (on the order of a few nm), it was found that the sample exposed to the saline solution had a slightly thicker oxide layer and more signifi cantly, a phosphate layer very near the surface suggesting toxic metal components are well contained within the sample. These are considerable indications of a biocompatible device.

Research Organization:
DOESC (USDOE Office of Science (SC) (United States))
Sponsoring Organization:
USDOE Office of Science (SC)
OSTI ID:
1052084
Journal Information:
Journal of Undergraduate Research, Vol. 8
Country of Publication:
United States
Language:
English

Similar Records

Corrosion resistance, chemistry, and mechanical aspects of Nitinol surfaces formed in hydrogen peroxide solutions
Journal Article · Tue Jun 12 00:00:00 EDT 2012 · Journal of Biomedical Materials Research · OSTI ID:1052084

Critical overview of Nitinol surfaces and their modifications for medical applications
Journal Article · Wed Feb 06 00:00:00 EST 2008 · ActaBiomaterialia · OSTI ID:1052084

Engineering Titanium for Improved Biological Response
Technical Report · Wed Jan 23 00:00:00 EST 2002 · OSTI ID:1052084

Related Subjects