Title: CC Cryostat Vacuum Pumping Efficiency

Technical Report ·
DOI:https://doi.org/10.2172/1030703· OSTI ID:1030703

This report calculates the effect of the conductances of the pumping lines on the pumping speeds of the vacuum pumps being used to pump the inner vessel, and annular space, vacuum tight during the CC Cryostat testing. Effective pumping speeds were calculated for various values of pressure via the above stated formulas (see calculations). Conductances of valves, elbows, and tees were calculated with the help of ref. 1, and the volumes of the inner vessel and annular space were calculated wtth the aid of ref. 2. The major results of these calculations follow. The attached graphs show the effective pumping speed vs. pressure, as well as the pressure vs. pumpdown time for both the inner vessel pumpdown and the annular space pumpdown. Many intervals of pressure were chosen in order to give a complete picture of the effects on the pumping speed, and pumpdown time. An important rule of thumb is that the effective pumping speed be close to the intrinsic pump speed in the region of interest for the best efficiency. With an infinite conductance, these quantities become equal. In the case of the annular space pumpdown. the effective pumping speed is within approximately 15% of the intrinsic pump speed for pressures down to 6000{mu}, and at that point, the conductance effects slow the system down and provide a 49% difference at 1000{mu}. To Improve these numbers, the line length must be shortened, or even better, the line diameter increased. As far as pumpdown time is concerned, the actual pumpdown was completed in a reasonable amount of time, and would have been even better if the pumping system did not lag due to a leaK and its repair time. The inner vessel pumpdown shows a pumpdown time near 1 hour to reach 100{mu} pressure. The effective pumping speed matches to within 10% of the intrinsic pump speed to 6000{mu}, and then the conductance effects become more important and provide a 41% difference at 1000{mu}. The actual pumpdown was completed in a reasonable amount of time which demonstrates that the pump used was in fact adequate for this application.

Research Organization:
Fermi National Accelerator Lab. (FNAL), Batavia, IL (United States)
Sponsoring Organization:
USDOE Office of Science (SC)
DOE Contract Number:
AC02-07CH11359
OSTI ID:
1030703
Report Number(s):
FERMILAB-D0-EN-117; TRN: US201201%%353
Country of Publication:
United States
Language:
English

Similar Records

CC Cryostat Cooldown Test Results
Technical Report · Fri Oct 16 00:00:00 EDT 1987 · OSTI ID:1030703

D0 Solenoid Upgrade Project: Vacuum Pumping Calculations for the D0 Solenoid
Technical Report · Mon Aug 02 00:00:00 EDT 1993 · OSTI ID:1030703

CC Pressure Test
Technical Report · Thu Jul 12 00:00:00 EDT 1990 · OSTI ID:1030703