Title: A high resolution x-ray fluorescence spectrometer for near edge absorption studies

Conference ·
OSTI ID:10109958

A high resolution fluorescence spectrometer using a Johann geometry in a back scattering arrangement was developed. The spectrometer, with a resolution of 0.3 eV at 6.5 keV, combined with an incident beam, with a resolution of 0.7 eV, form the basis of a high resolution instrument for measuring x-ray absorption spectra. The advantages of the instrument are illustrated with the near edge absorption spectrum of dysprosium nitrate. 10 refs., 4 figs.

Research Organization:
Brookhaven National Lab., Upton, NY (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC02-76CH00016
OSTI ID:
10109958
Report Number(s):
BNL-46809; CONF-910730-28; ON: DE92004991
Resource Relation:
Conference: 4. international conference on synchrotron radiation instrumentation,Chester (United Kingdom),15-19 Jul 1991; Other Information: PBD: [1991]