SPECT assay of radiolabeled monoclonal antibodies. Progress report, September 1, 1992--August 24, 1993
The overall goal of this project is to improve the effectiveness of single photon emission computed tomography (SPECT) to image and quantify radiolabeled monoclonal antibodies. During the past year, we have made significant progress toward this goal, and this report summarizes that work. Our efforts have been mainly directed along three fronts. First, we have developed and tested new reconstruction methods including three-dimensional iterative algorithms that model non-uniform attenuation and distance-dependent detector response. Both fan beam and parallel beam collimator geometries have been modeled and novel ways of improving the efficiency of the computationally intensive methods have been introduced. Second, an ultra-high resolution, small field-of-view pinhole collimator has been constructed and evaluated. Reconstructed spatial resolution of 1 to 3 mm (FWHM) has been achieved in phantom scans with a useful field-of-view of 9 to 10 cm. Finally, we have investigated the ability of SPECT to image and quantify astatine-211 distributions. Reconstructed images of phantom data demonstrated quantitative accuracy to within 10% with proper attenuation and scatter compensation.
- Research Organization:
- Duke Univ. Medical Center, Durham, NC (United States). Dept. of Radiology
- Sponsoring Organization:
- USDOE, Washington, DC (United States)
- DOE Contract Number:
- FG05-89ER60894
- OSTI ID:
- 10103355
- Report Number(s):
- DOE/ER/60894-5; ON: DE94001952; BR: KD0600000/KP0601000
- Resource Relation:
- Other Information: PBD: 20 Aug 1993
- Country of Publication:
- United States
- Language:
- English
Similar Records
SPECT assay of radiolabeled monoclonal antibodies. Final performance report, March 1992--November 1995
SPECT assay of radiolabeled monoclonal antibodies. Third yearly progress report, September 1991--February 1992
Related Subjects
MONOCLONAL ANTIBODIES
SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY
IMAGE PROCESSING
ALGORITHMS
DESIGN
PROGRESS REPORT
RESOLUTION
ASTATINE 211
THREE-DIMENSIONAL CALCULATIONS
ITERATIVE METHODS
COLLIMATORS
RADIOIMMUNOSCINTIGRAPHY
550601
UNSEALED RADIONUCLIDES IN DIAGNOSTICS