skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Microfabricated thermal conductivity detector for the micro-ChemLab.

Journal Article · · Proposed for publication in Sensors and Actuators B.
OSTI ID:972450

This work describes the design, computational prototyping, fabrication, and characterization of a microfabricated thermal conductivity detector ({mu}TCD) to analyze the effluent from a micro-gas chromatograph column ({mu}GC) and to complement the detection efficacy of a surface acoustic wave detector in the micro-ChemLab{trademark} system. To maximize the detection sensitivity, we designed a four-filament Wheatstone bridge circuit where the resistors are suspended by a thin silicon nitride membrane in pyramidal or trapezoidal shaped flow cells. The geometry optimization was carried out by simulation of the heat transfer in the devices, utilizing a boundary element algorithm. Within microfabrication constraints, we determined and fabricated nine sensitivity-optimized geometries of the {mu}TCD. The nine optimal geometries were tested with two different flow patterns. We demonstrated that the perpendicular flow, where the gas directly impinged upon the membrane, yielded a sensitivity that is three times greater than the parallel flow, where the gas passed over the membrane. The functionality of the {mu}TCD was validated with the theoretical prediction and showed a consistent linear response to effluent concentrations, with a detection sensitivity of 1 ppm, utilizing less than 1 W of power.

Research Organization:
Sandia National Laboratories (SNL), Albuquerque, NM, and Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC04-94AL85000
OSTI ID:
972450
Report Number(s):
SAND2005-4926J; TRN: US201006%%86
Journal Information:
Proposed for publication in Sensors and Actuators B., Journal Name: Proposed for publication in Sensors and Actuators B.
Country of Publication:
United States
Language: