skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Incorporation of a Flared Inlet Capillary Tube on a Fourier Transform Ion Cyclotron Resonance Mass Spectrometer.

Journal Article · · Journal of the American Society for Mass Spectrometry, 17(6):772-779

Atmospheric pressure ion sources such as electrospray ionization (ESI) and atmospheric pressure matrix assisted laser desorption ionization (AP-MALDI) are widely used with mass spectrometry for proteomics studies. Other newly developed atmospheric ion sources include desorption electrospray ionization (DESI). For these ionization methods, analyte ions created at atmospheric pressure are transferred to the high vacuum region for mass analysis after several stages of differential pumping. It has been reported that overall charge transmission between the ion source and the first vacuum stage is primarily dependent upon the proximity of the emitter and gas conductance of the interface inlet. We therefore developed an atmospheric pressure interface using flared inlet tubes. This report highlights our results obtained by coupling the customized flared inlet tubes with a Fourier transfer ion cyclotron resonance mass spectrometer (FT-ICRMS). We have also investigated the new interface with different types of atmospheric pressure ionization methods. For most of the ionization methods we investigated, such as ESI and DESI, increased ion current transmitted from the atmospheric pressure ion source to the first stage vacuum system was observed with the use of our enhanced ion inlet designs. The ion intensity that was detected with the flared inlet tube on a FT-ICRMS was also observed to increase {approx} 2-5 fold using ESI or DESI with the flared tube inlet. Moreover, increased spray tip positional tolerance was observed with implementation of the flared inlet tube. We also include our preliminary results obtained by coupling APMALDI with flared inlet tube in this paper. For AP-MALDI, the measured ion current transferred through the flared inlet tube was about 3 times larger than the ion current through the control non-flared inlet tube.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
892880
Report Number(s):
PNNL-SA-48140; TRN: US200623%%517
Journal Information:
Journal of the American Society for Mass Spectrometry, 17(6):772-779, Journal Name: Journal of the American Society for Mass Spectrometry, 17(6):772-779
Country of Publication:
United States
Language:
English