skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Performance of a PSPMT based detector for scintimammography

Journal Article · · Phys. Med. Biol.
OSTI ID:804874

In breast scintigraphy, compact detectors with high intrinsic spatial resolution and small inactive peripheries can provide improvements in extrinsic spatial resolution, efficiency and contrast for small lesions relative to larger conventional cameras. We are developing a pixelated small field-of-view gamma camera for scintimammography. Extensive measurements of the imaging properties of a prototype system have been made, including spatial resolution, sensitivity, uniformity of response, geometric linearity and energy resolution. An anthropomorphic torso phantom providing a realistic breast exit gamma spectrum has been used in a qualitative study of lesion detectability. A new type of breast imaging system that combines scintimammography and digital mammography in a single upright unit has also been developed. The system provides automatic co-registration between the scintigram and the digital mammogram, obtained with the breast in a single configuration.

Research Organization:
Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)
Sponsoring Organization:
USDOE Office of Energy Research (ER) (US)
DOE Contract Number:
AC05-84ER40150
OSTI ID:
804874
Report Number(s):
JLAB-PHY-00-56; DOE/ER/40150-2391; TRN: US0205574
Journal Information:
Phys. Med. Biol., Vol. 45, Issue 3; Other Information: Phys. Med. Biol. 45 No 3 (March 2000) 781-800; PBD: 1 Mar 2000
Country of Publication:
United States
Language:
English