Title: An economical x-ray area detector based on an intensified CCD unit

Journal Article · · Review of Scientific Instruments; (United States)
DOI:https://doi.org/10.1063/1.1144262· OSTI ID:7185385
; ; ; ;  [1]
  1. Chemistry Department, State University of New York at Stony Brook, Stony Brook, New York 11794-3400 (United States)

An economical charge-coupled device (CCD) based area detector for x-ray experiments was developed and assembled. Although low in cost, careful evaluation shows that this detector consisting of an optical fiber taper, an image intensifier, a relay lens, and a cooled CCD unit has good overall features and is comparable with other reported CCD based x-ray area detectors. The detector is suitable for small-angle x-ray scattering and wide-angle x-ray diffraction experiments employing either synchrotron or conventional x-ray sources. It also has several options which we could utilize under different experimental conditions. The test result, its limitations, and possible improvements are discussed.

DOE Contract Number:
FG05-89ER75515; FG02-86ER45237
OSTI ID:
7185385
Journal Information:
Review of Scientific Instruments; (United States), Vol. 64:2; ISSN 0034-6748
Country of Publication:
United States
Language:
English