skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: H/sub. cap alpha. / laser fluorescence diagnostic on the Tara Tandem Mirror experiment

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1139691· OSTI ID:7123675

A laser fluorescence diagnostic has been used for measuring the neutral hydrogen density in the central cell of the Tara thermal barrier tandem mirror. Experiments have been performed using laser-induced, resonance fluorescence detection of H/sub ..cap alpha../ (6563-A) radiation. Measurements were made at a number of radial positions with 1-cm resolution, from the magnetic axis to near the plasma limiter. Stray laser light contributions to the signal were eliminated with a double-pulse technique. For comparison, the chord-averaged plasma H/sub ..cap alpha../ radiation was analyzed under the identical conditions for which laser fluorescence data were taken.

Research Organization:
Plasma Fusion Center, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139
OSTI ID:
7123675
Journal Information:
Rev. Sci. Instrum.; (United States), Vol. 59:8
Country of Publication:
United States
Language:
English