skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Physical characteristics of TTV03, a new high spiral resolution time-of-flight position tomograph

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA)
OSTI ID:6587707
; ; ; ;  [1]; ; ; ; ;  [2]
  1. Service Hospitalier Frederic Joliot, CEA, 91406 Orsay (FR)
  2. Div. LETI, CEA, 38000 Grenoble (FR)

The authors present TTV03, a new high spatial resolution TOF-PET system. The machine consists of four 89 cm diameter rings made of 7 mm wide, 18 mm high and 45 mm deep BaF{sub 2} crystals (324 per ring) individually coupled to 11 mm diameter phototubes. The transaxial resolution (FWHM) in the reconstructed image is 5.7 mm at center, 5.7 mm (tangential) and 6.5 mm (radial) at 10 cm. The axial resolution is 8.8 mm (direct plane) and 6.7 mm (crossplane) at center, 10.3 mm (direct) and 11.2 mm (cross) at 10 cm from center. With a 150 keV threshold, the time-of-flight resolution is 750 psec (FWHM), and, with a 4.18 nsec coincidence window, the sensitivity (true evts/.sec per {mu}Ci/cc) is 9800 (direct plane) and 14300 (crossplane).

OSTI ID:
6587707
Report Number(s):
CONF-900143-; CODEN: IETNA; TRN: 90-031692
Journal Information:
IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA), Vol. 37:2; Conference: Institute for Electronic and Electrical Engineers (IEEE) nuclear science symposium, San Francisco, CA (USA), 15-19 Jan 1990; ISSN 0018-9499
Country of Publication:
United States
Language:
English