skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Sampling of airborne polycyclic aromatic hydrocarbons

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00140a025· OSTI ID:6418622

Limitations of NIOSH sampling method P and CAM 183 were defined for airborne standard mixtures of polycyclic aromatic hydrocarbons (PAH) generated as vapors in a flow-through apparatus. The PAH fell into three categories: those that were too volatile to be collected by the NIOSH filtration method at normal ambient temperatures and were best sampled with Tenax or XAD-2 sorbent (i.e., indane, naphthalene, biphenyl, acenaphthene, fluorene, 9,10-dihydrophenanthrene, phenathrene, and anthracene); those that were quantitatively collected by filters, even after a brief airborne residence time (i.e., benz(a)anthracene, chrysene, benzo(a)pyrene, dibenz(a,h)anthracene, and benzo(ghi)perylene); and those that partitioned between filter and sorbent (i.e., fluoranthene and pyrene). A combination glass fiber/silver membrane filter backed by two sorbent tubes in series gave overall mean recoveries of 94-96% for the 15 PAH studied at total concentrations of, nominally, 0.2 and 0.02 mg/m/sup 3/. Individual PAH concentrations were 0.03-0.05 and 0.003-0.005 mg/m/sup 3/, respectively.

Research Organization:
Health and Welfare Canada, Ottawa, Ontario
OSTI ID:
6418622
Journal Information:
Anal. Chem.; (United States), Vol. 59:13
Country of Publication:
United States
Language:
English