Title: Hazardous sound levels produced by extracorporeal shock wave lithotripsy

Journal Article · · J. Urol.; (United States)
OSTI ID:6325978

Sound emitted from the Dornier system GmbH lithotriptor was found to be of sufficient intensity to warrant concern about noise-induced sensorineural hearing loss. The patients were exposed to impulses of 112 dB. peak sound pressure level. Operating room personnel were exposed to sounds of less intensity, although the number of impulses they were exposed to was much greater, thereby increasing the risk of hearing loss. Hearing protection is recommended for patients and operating room personnel.